beautypg.com

6 appendix, 1 technical data, 1 conductivity measurement – Metrohm 792 Basic IC User Manual

Page 168: 2 conductivity detector

background image

6.1 Technical data

792 Basic IC

159

6 Appendix

6.1 Technical

data

6.1.1 Conductivity

measurement

Measurement range

0}1000 PS/cm (resolution: 0.56 nS/cm)

Maximum error

r 10 PS/cm and
r 1 % of measurement value (k = 16.7/cm)

Linearity

Deviations < r 5 PS/cm

Noise

typ. 3 nS/cm

Drift (electronic)

typ. < 10 nS/cm / h

Temperature dependence

typ. < 40 nS/cm / °C

Reserve range

> 33 % (k = 16.7/cm)

Sampling rate

10 measurements/s (fixed)

6.1.2 Conductivity

detector

Construction

Thermostatted conductivity detector with

2 ring-shaped steel electrodes

Measurement principle

Alternating current measurement with 1 kHz

frequency and ca. 1.7 V amplitude (peak to

peak).

Cell volume

1.5 PL

Cell constant

approx. 17 /cm (the exact value is printed on the

detector)

Maximum back pressure

for measuring cell

5.0 MPa (50 bar)

Thermostatting

Connectable dynamic control to operating

temperature

Operating temperature

40 °C

Max. temperature deviation

r 2.5°C

Heating time

t 30 min

Temperature stability

d 0.01°C at constant ambient temperature

Connection for detector block Dsub 15 pin (female)