Metrohm 792 Basic IC User Manual

Page 112

background image

4.4 Methods

792 Basic IC

103

Calibration points table
This table contains the basic information used to construct the calibra-

tion curve:

Level

Number of

Calibration level

.

Conc.

Concentration of the current component in the

calibration sample. It is taken from the concen-

tration table.

Area (or Height)

Peak height or area of the current component,

depending on calibration base.

File

File name of calibration chromatogram that

stores data on the given calibration level.

Used

Information whether the calibration level is used

for calculation of the calibration curve or not.

Exclude the selected calibration point from the

list and recalculate calibration coefficients for

the current component. Repeated pressing in-

cludes the point again. You can exclude points

that drop out of the calibration curve watching

for RSD values.

Component information

Component

Allows to select the current component from the

list. It is possible also to scroll the list of com-

ponents by mouse, using the special arrow but-

tons on the right.

Retention time

Retention time of the selected component

(read-only).

Concentration

Concentration of the selected component

(read-only).

Calibration parameters

Calibration method

Method that is used for calibration procedure.

There are three basic methods of calibration:

External standard calibration

Absolute calibration. It is the basic calibration

procedure in ion chromatography.

Internal standard calibration

Relative calibration.

Tabulated calibration

Relative response factor calibration. It is a sim-

plified method of external standard calibration.

Local

If this checkbox is marked then the correspond-

ing parameter is valid for the current compo-

nent only. Otherwise, this parameter is global

(i.e. default for all other components).