Replacing the suppressor
The suppressor in the suppressor block may have to be replaced in the
following cases:
x Irremediable loss of suppressor capacity (reduced
phosphate sensitivity and/or strong rise in baseline)
x Irremediable blockage of the suppressor (the suppres-
sor can no longer deliver solutions)
Both the 6.2832.000 Suppressor rotor and the 6.2832.010 Connection
piece with the input and output leads can be replaced. To replace these
components proceed as follows (see Fig. 19):
1 Disconnect suppressor from IC system
x Disconnect all input and output leads of the suppressor from
IC system and peristaltic pump.
2 Dismantle suppressor
x Unscrew nut 102
102
102
102 from suppressor holder 105
105
105
105.
x Pull out connection piece 103
103
103
103 and suppressor rotor 104
104
104
104 from
suppressor holder 105
105
105
105 (the connection piece and the rotor
normally stick together).
x Loosen connection piece 103
103
103
103 from suppressor rotor 104
104
104
104.