beautypg.com

9 replacing the suppressor – Metrohm 792 Basic IC User Manual

Page 162

background image

5.2 Maintenance and servicing

792 Basic IC

153

105

105

105

105

104

104

104

104

103

103

103

103

102

102

102

102

Fig. 19: Assembling the suppressor

102

102

102

102

Screw nut

104

104

104

104

Suppressor rotor (6.2832.000)

103

103

103

103

Connection piece (6.2832.010)

with input and output leads

105

105

105

105

Suppressor holder






5.2.9

Replacing the suppressor

The suppressor in the suppressor block may have to be replaced in the

following cases:

x Irremediable loss of suppressor capacity (reduced

phosphate sensitivity and/or strong rise in baseline)

x Irremediable blockage of the suppressor (the suppres-

sor can no longer deliver solutions)

Both the 6.2832.000 Suppressor rotor and the 6.2832.010 Connection

piece with the input and output leads can be replaced. To replace these

components proceed as follows (see Fig. 19):

1 Disconnect suppressor from IC system

x Disconnect all input and output leads of the suppressor from

IC system and peristaltic pump.

2 Dismantle suppressor

x Unscrew nut 102

102

102

102 from suppressor holder 105

105

105

105.

x Pull out connection piece 103

103

103

103 and suppressor rotor 104

104

104

104 from

suppressor holder 105

105

105

105 (the connection piece and the rotor

normally stick together).

x Loosen connection piece 103

103

103

103 from suppressor rotor 104

104

104

104.