beautypg.com

Metrohm 792 Basic IC User Manual

Page 85

background image

4 Operation

792 Basic IC

76

Configuration

The

Configuration

tab in the system settings window contains configura-

tion settings for the 792 Basic IC.

Method

Selection of the method with or without sup-

pressor:

With MSM

method with suppressor

Without MSM

method without suppressor

Suppressor

Suppressor

module:

Autostep with

Automatic switching to the next position each

time the injection valve is switched either to the

Fill

or to the

Inject

position.

Use peristaltic pump

If this option is disabled, the peristaltic pump is

not switched on automatically at the start of a

determination or with

Startup hardware

.

Polarity

Selection of the polarity of the output signal:

+

positive polarity (for anions)

negative polarity (for cations)




Hardware settings

Selection of the

Hardware

menu item using the right mouse button

opens the hardware settings window consisting of the two tabs

Hard-

ware

and

Alarm stops

.

Hardware
This tab of the hardware settings window defines general parameters

which are set automatically at power on of the instrument.