6 batch reprocessing, 1 batch reprocessing queue file handling – Metrohm 792 Basic IC User Manual
Page 141
4 Operation
792 Basic IC
132
4.6 Batch
reprocessing
Reprocessing is understood to be the subsequent reprocessing of a
series of chromatograms which have been loaded in a
queue (Batch
reprocessing queue). For reprocessing according to a selected
method the settings for calibration, integration, passport, appearance
and report can be altered at will.
A batch reprocessing queue is stored in a batch reprocessing queue
file
*.bar
in the
Data
directory.
4.6.1
Batch reprocessing queue file handling
Open batch reprocessing queue
792 BASIC IC / File / Open / Batch reprocessing
Load an existing batch reprocessing file (
*.bar
) from the
Data
directory
and open the
Reprocess
window.
792 BASIC IC / File / Open / Last batch
Load the last opened batch reprocessing file (
*.bar
) from the
Data
direc-
tory and open the
Reprocess
window.
Create new batch reprocessing queue
792 BASIC IC / File / Open / Chromatogram
For creation of a new batch reprocessing queue, open the
Chroma-
togram open
window. Select the desired chromatograms (
*.chw
) and
click the
button. Enter the name of the new batch reprocess-
ing file and press
. The selected chromatograms are then added
to this batch reprocessing queue.
If the reprocessing includes reintegration and/or recalibration, only
chromatograms recorded with the same method should be loaded
into the batch reprocessing queue.
Save batch reprocessing queue
QUEUE EDITOR / File / Save
Save the batch reprocessing queue in a batch reprocessing queue file
(
*.bar
) in the working directory. The batch reprocessing queue editor
window remains open.
QUEUE EDITOR / File / Save & exit
Save the batch reprocessing queue in a batch reprocessing queue file
(
*.bar
) in the working directory and close the batch reprocessing editor
window.