3 precolumn glass cartridges with cartridge holder – Metrohm 792 Basic IC User Manual

Page 39

background image

2 Installation

792 Basic IC

30

2.7.3

Precolumn glass cartridges with cartridge holder

The precolumn glass cartridge METROSEP Anion Dual 1 (6.1006.030)

is inserted in the 6.2828.110 Precolumn cartridge holder as follows (see

Fig. 11):

1 Insert cartridge

x Remove end fittings 62

62

62

62 from sleeve 63

63

63

63.

x Insert precolumn cartridge 64

64

64

64 in sleeve 63

63

63

63 (mark flow direc-

tion of the column cartridge on sleeve, the column cartridge

should always be operated in the same flow direction).

x Screw both end fittings 62

62

62

62 loosely to sleeve 63

63

63

63 by hand.

x Tighten both end fittings 62

62

62

62 first by hand, then firmly using a

screwdriver (½ ¼ turn).

2 Connect precolumn

x Provide the column connection capillary 20

20

20

20 mounted on the

injection valve with a PEEK compression fitting 46

46

46

46 (see sec-

tion 2.3.4) and screw it tightly onto end fitting 62

62

62

62 on the inlet

side of the precolumn.

x Screw connection piece 65

65

65

65 to end fitting 62

62

62

62 on the other end

of the precolumn.

3 Rinse precolumn

x Place a beaker beneath the outlet capillary of the precolumn.
x Open software window for manual system control.
x If necessary, modify

Flow rate

to the value suited for the

inserted separating column and click on

to

send this value to the 792 Basic IC.

x Switch on high-pressure pump (

IC pump

) by clicking

and

rinse precolumn with eluent for ca. 10 min.

x Switch off high-pressure pump by clicking

.