Metrohm 792 Basic IC User Manual
Page 114
4.4 Methods
792 Basic IC
105
Indexes
Fill
Index
field for components with zero index.
Coefficients
Recalculate all calibration coefficients for all
components (performs re-calibration).
Coef. (balance)
Recalculate coefficient of the universal compo-
nent. The aim of this action is to adjust the re-
sponse factor of this component so, that the to-
tal amount of the injected components
becomes equal to the declared value. This de-
clared summary amount is entered into the
Concentration
field of the universal component
for
This run
level.
Load and save calibration data
792 BASIC IC / Method / Calibration / Load from method
Load the calibration from the method. This option is used to update the
calibration for the current run by the one taken from the method used
for acquisition.
792 BASIC IC / Method / Calibration / Save to method
Save the current calibration to the method. This option is used to trans-
fer the updated calibration from the run to the current method (i.e. the
method associated with the current chromatogram window).
Import and export calibration data
792 BASIC IC / Method / Calibration / Import calibration
This option imports a calibration from a calibration file
*.cal
to the cur-
rent method or chromatogram. It is used to transfer the updated cali-
bration from chromatogram-to-chromatogram or from method-to-
method. Different methods can be processed in this case using the
same calibration.
792 BASIC IC / Method / Calibration / Export calibration
This option exports a calibration to a in calibration file
*.cal
. It is useful to
transfer a calibration from chromatogram-to-chromatogram for chroma-
tograms that were obtained by different methods.
Put out calibration curves
COMPONENT / Copy to clipboard
This option allows to copy the calibration curve of the selected compo-
nent to the clipboard so that it is available for other Windows applica-
tions, such as WinWord, Excel, etc.
COMPONENT / Print/Preview / Preview this
Display calibration curve print preview of the selected component.
COMPONENT / Print/Preview / Preview all
Display calibration curve print preview of all components.
COMPONENT / Print/Preview / Print this
Print selected calibration curve of the selected component.
COMPONENT / Print/Preview / Print all
Print calibration curves of all components.