4 Operation
792 Basic IC
78
Event
Events for alarm stop:
Pump stop
Pump stopped because pressure limits are ex-
ceeded.
Hardware error
Hardware error detected at the 792 Basic IC
(high-pressure pump, injection valve, or sup-
pressor not working correctly).
Power on
Temporary power failure at the 792 Basic IC.
4.3.7
System state window
The
SYSTEM STATE
window is automatically opened if a system is
opened. It shows status and error messages for this system. Messages
concerning the 792 Basic IC hardware are followed by
[792 Basic IC]
,
messages concerning the loaded system are followed by
["folder"]
(name of subfolder containing the system file).
Status messages
Checking on-line
Checking connection between PC and 792 Basic IC.
On-line
Connection between PC and 792 Basic IC ok.
UploadStartupValues
System startup values have been loaded to the 792
Basic IC.
Initialization
Hardware or system initialization.
Ready
System is ready for starting a new determination.
Starting
Starting program or chromatogram data acquisition.
Running
Running program or chromatogram data acquisition.
Running program
(xxx min left)
Running time program
(in brackets: time left).
Waiting for INJECT
Waiting for "INJECT" to start program and/or
chromatogram data acquisition.
INJECT done
Injection valve has been switched to the "INJECT"
position.
Finished
Determination has been finished.
SHUTDOWN
792 Basic IC hardware is shutdown