4 Operation

792 Basic IC

78

Event

Events for alarm stop:

Pump stop

Pump stopped because pressure limits are ex-

ceeded.

Hardware error

Hardware error detected at the 792 Basic IC

(high-pressure pump, injection valve, or sup-

pressor not working correctly).

Power on

Temporary power failure at the 792 Basic IC.

4.3.7

System state window

The

SYSTEM STATE

window is automatically opened if a system is

opened. It shows status and error messages for this system. Messages

concerning the 792 Basic IC hardware are followed by

[792 Basic IC]

,

messages concerning the loaded system are followed by

["folder"]

(name of subfolder containing the system file).

Status messages

Checking on-line

Checking connection between PC and 792 Basic IC.

On-line

Connection between PC and 792 Basic IC ok.

UploadStartupValues

System startup values have been loaded to the 792
Basic IC.

Initialization

Hardware or system initialization.

Ready

System is ready for starting a new determination.

Starting

Starting program or chromatogram data acquisition.

Running

Running program or chromatogram data acquisition.

Running program
(xxx min left)

Running time program
(in brackets: time left).

Waiting for INJECT

Waiting for "INJECT" to start program and/or
chromatogram data acquisition.

INJECT done

Injection valve has been switched to the "INJECT"
position.

Finished

Determination has been finished.

SHUTDOWN

792 Basic IC hardware is shutdown