Metrohm 792 Basic IC User Manual
Page 82
4.3 Systems
792 Basic IC
73
Flow, mL/min
Startup value for flow rate of the high-pressure
pump.
Entry range:
0.20
...
2.50 mL/min
Pressure max, MPa
Startup value for maximum pressure limit for
high-pressure pump.
Entry range:
0.0
...
25.0 MPa
Pressure min, MPa
Startup value for minimum pressure limit for
high-pressure pump.
Entry range:
0.0
...
25.0
MPa
Instrument control for connected system
A double-click on the instrument icon or selection of the
Open
menu
item using the right mouse button opens the system settings window.
For a connected system, the
Control
tab is displayed. It allows manual
control of the 792 Basic IC functions and setting of startup values to be
sent to the instrument. This tab shows the current measurement values
for conductivity and pressure.
Conductivity, PPPPS/cm
Live display of measured conductivity.
Pressure, MPa
Live display of measured pressure.
The color settings for this two fields can be
changed by clicking the fields with the right
mouse button and selecting the menu item
Choose color / ...
.
Actual values
Display of actual values
Flow, mL/min
Display of flow rate of the high-pressure pump.
Pressure max, MPa
Display of maximum pressure limit for high-
pressure pump.
Pressure min, MPa
Display of minimum pressure limit for high-
pressure pump.