beautypg.com

Metrohm 792 Basic IC User Manual

Page 82

background image

4.3 Systems

792 Basic IC

73

Flow, mL/min

Startup value for flow rate of the high-pressure

pump.

Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum pressure limit for

high-pressure pump.

Entry range:

0.0

...

25.0 MPa

Pressure min, MPa

Startup value for minimum pressure limit for

high-pressure pump.

Entry range:

0.0

...

25.0

MPa

Instrument control for connected system

A double-click on the instrument icon or selection of the

Open

menu

item using the right mouse button opens the system settings window.

For a connected system, the

Control

tab is displayed. It allows manual

control of the 792 Basic IC functions and setting of startup values to be

sent to the instrument. This tab shows the current measurement values

for conductivity and pressure.

Conductivity, PPPPS/cm

Live display of measured conductivity.

Pressure, MPa

Live display of measured pressure.
The color settings for this two fields can be

changed by clicking the fields with the right

mouse button and selecting the menu item

Choose color / ...

.

Actual values

Display of actual values

Flow, mL/min

Display of flow rate of the high-pressure pump.

Pressure max, MPa

Display of maximum pressure limit for high-

pressure pump.

Pressure min, MPa

Display of minimum pressure limit for high-

pressure pump.