2.4 Mains connection
792 Basic IC
17
2.4 Mains
connection
Follow the instructions below for connecting to the power supply. If
the instrument is operated with a mains voltage set wrongly and/or
wrong mains fuse, there is a danger of fire!
2.4.1
Setting the mains voltage
Before switching on the 792 Basic IC for the first time, check that the
mains voltage set on the instrument (see Fig. 8) matches the local
mains voltage. If this is not the case, you must reset the mains voltage
on the instrument as follows:
1 Disconnect mains cable
Disconnect mains cable from mains connection plug 12
12
12
12 of the
792 Basic IC.
2 Remove fuse holder
Using a screwdriver, loosen fuse holder 13
13
13
13 below the mains
connection plug 12
12
12
12 and take out completely.
3 Check and change fuse if necessary
Carefully take the fuse installed for the desired mains voltage out
of fuse holder 13
13
13
13 and check its specifications (the position of the
fuse in the fuse holder is marked by the white arrow imprinted
next to the mains voltage range):
100}
}
}
}120 V 1.0 A (slow-blow) Metrohm No. U.600.0016
220}
}
}
}240 V 0.5 A (slow-blow) Metrohm No. U.600.0013
4 Insert fuse
Change fuse if necessary and reinsert in fuse holder 13
13
13
13.
5 Install fuse holder
Depending on the desired mains voltage, insert fuse holder 13
13
13
13 in
the 792 Basic IC so that the corresponding mains voltage range
can be read normally and the adjacent white arrow points to the
white bar imprinted below the fuse holder (see Fig. 8).