beautypg.com

Metrohm 792 Basic IC User Manual

Page 130

background image

4.5 Chromatograms

792 Basic IC

121

Export chromatogram

792 BASIC IC / File / Export / AIA file

Export the selected chromatogram in the AIA format (Analytical Instru-

ment Association) as CDF file (

*.cdf

).

792 BASIC IC / File / Export / Raw data to txt

Export chromatographic raw data into an ASCII text file (

*.txt

). The fol-

lowing window appears:

Represent raw data as

Output of raw data:

Default units (float)

Y values are exported in default units for the

channel (PS/cm).

ADC response (integer)

Y values are exported in bits.

Include retention time

In seconds (float)

X values are exported in seconds.

As point number

X values are exported as point number.