beautypg.com

5 suppressor module, 6 connections – Metrohm 792 Basic IC User Manual

Page 151

background image

5 Notes – Maintenance – Faults

792 Basic IC

142

5.1.5 Suppressor

module

Protection

To avoid contamination of the suppressor module by foreign particles

or bacterial growth, the

two 6.2821.100 Filter units PEEK (see sec-

tion 2.3.5) supplied must be installed between peristaltic pump and inlet

capillaries of the suppressor module (see section 2.8.2).

Operation

The

Metrohm Suppressor Module MSM comprises a total of 3 sup-

pressor units which are in turn used for suppression, regenerated with

sulfuric acid and rinsed with water. To record every new chromatogram

under comparable conditions, work is normally carried out with freshly

regenerated suppressor. Switching takes place automatically together

with the valve switching.

For operation of the suppressor module, a

two-channel peristaltic

pump is needed which conveys the regeneration solution (normally

20 mmol/L H

2

SO

4

) and the rinsing solution (normally

dist. H

2

O) to the

suppressor units (flow rate of 0.5 mL/min).

The suppressor units must never be regenerated with H

2

SO

4

in the

same flow direction used for the eluent. You should thus always install

the inlet and outlet capillaries as described in section 2.8.4 according

to the scheme shown in Fig. 15.

The suppressor module must never be switched in the dry state as

there is a danger of blocking. Before every switching operation of the

suppressor module, the three suppressor units must have been

rinsed for at least ½ h with eluent, regeneration and rinsing solution.

The capacity of the suppressor units is exhausted after approx. 2 h.

With longer interruptions between the individual measurements, it is

recommended that the

Prep MSM.smt

system is started during the

pauses; this automatically switches the suppressor module further

every 20 min.

Maintenance

In the event of reduced capacity or high counter-pressure the suppres-

sor module must be regenerated (section 5.2.7), cleaned (section 5.2.8)

or exchanged (section 5.2.9).

5.1.6 Connections

All connections between injector, column and detector must be as short

as possible, have a low dead volume and be absolutely tight. The PEEK

capillary after the detector block must be free from constriction (the

measuring cell is tested to 5 MPa = 50 bar back pressure).