beautypg.com

Metrohm 792 Basic IC User Manual

Page 91

background image

4 Operation

792 Basic IC

82

To reset the

Run

number press [Ctrl] [F8].

The

Analysis number

window is opened where

the run number for the next run can be

modified to the desired value.

Batch

X/Y: total number of started sample queues /

current analysis number in the sample queue

(read-only).


Sample

Tab

Sample

of the passport with sample information.

Info 1

First sample description (max. 256 characters).

Info 2

Second sample description (max. 256 charac-

ters).

Volume

Injected volume in PL.

Dilution

Dilution of the sample.

Vial number

Autosampler vial position to take sample from.

Amount

Sample amount. If this value is different for the

calibration run (c) and the sample run (s), the

component concentrations of the sample are

calculated as follows:

C

s

= C

c

Amount

s

/

Amount

c

Internal standard amount

Concentration of the internal standard compo-

nent for relative concentration calculations.

Date/time when...

Date and time of sample collection (the default

values are equal to the date and time when the

chromatogram starts).