12 - 2

(4) Setting

Transmission speed, parity and stop bit should be set prior to transmission.

Set values before setting.

Transmission speed

: 300 bps

Parity

: None

Stop bit

: 1 bit

[Note] After setting, the set data in memory are backed up by built-in battery.

(5) Communication control

q

Method

Transmission of data with flow trasnmitter is made according to request command from host computer.

When a command is received from host computer, the data corresponding to the command is
transmitted from Portaflow to the host computer.

w

Commands

Commands with ASCII code are transmitted from host computer to flow transmitter.

The last code of 1 command is carriage return (0DH).

[Portaflow RS-232C Transmission Command Table]

Real time transmission

Command

DQ01 [CR]

Instantaneous flow (L/s or gal/s)

1.000E + 3

L/s

[CR] [CR]

DQ02 [CR]

Instantaneous flow (L/min or gal/min)

1.000E + 3

L/min

[CR] [CR]

DQ03 [CR]

Instantaneous flow (L/h or gal/h)

1.000E + 3

L/h

[CR] [CR]

DQ04 [CR]

Instantaneous flow (ML/d or Mgal/d)

1.000E + 3

ML/d

[CR] [CR]

DQ05 [CR]

Instantaneous flow (m

3

/s or ft

3

/s)

1.000E + 3

m3/s

[CR] [CR]

DQ06 [CR]

Instantaneous flow (m

3

/min or ft

3

/min) 1.000E + 3

m3/min

[CR] [CR]

DQ07 [CR]

Instantaneous flow (m

3

/h or ft

3

/h)

1.000E + 3

m2/h

[CR] [CR]

DQ08 [CR]

Instantaneous flow (Mm

3

/d or Mft

3

/d)

1.000E + 3

Mm3/d

[CR] [CR]

DQ09 [CR]

Instantaneous flow (BBL/s)

1.000E + 3

BBL/s

[CR] [CR]

DQ10 [CR]

Instantaneous flow (BBL/min)

1.000E + 3

BBL/min

[CR] [CR]

DQ11 [CR]

Instantaneous flow (BBL/h)

1.000E + 3

BBL/h

[CR] [CR]

DQ12 [CR]

Instantaneous flow (MBBL/d)

1.000E + 3

MBBL/d

[CR] [CR]

DV [CR] Instantaneous flow (m/s or ft/s)

32.00E + 1

m/s

[CR] [CR]

DI+ [CR]

Normal direction integration flow

+ TOTAL 1234567 mL

[CR] [CR]

DI- [CR]

Reverse direction integration flow

- TOTAL 0000123 mL

[CR] [CR]

DAO [CR]

Analog output

1.000E + 2 Analog Out

[CR] [CR]

DAI [CR]

Analog input

1.000E + 2 Analog In

[CR] [CR]

DC [CR]

Status

NORMAL

[CR] [CR]

(output of status comment of up to 20 characters)

DT [CR]

Time

95 - 01 - 05 12 : 00

[CR] [CR]

Data

Data format (example)

: space

Space for head of flow rate/velocity ( ) shows (-) at reverse flow.

BULLETIN F-68