8 - 2

q

q

q

q

q “Indicator”

Shows the intensity of ultrasonic receiving signal.
Check if 2 or more indicators are displayed.
If one or less indicator is displayed, raise the transmission voltage level as shown on page 6-12.
When the sensor is not connected, the indicator may be lighted by sensing noises. But, this is not
an error.

w

w

w

w

w “Status display”

Check if “NORMAL” is displayed. If the sensor is not connected, other messages may be displayed.
This is not an error.
In case other message is displayed after installing and connecting the sensor, take corrective actions
according to page 9-33, “System check function”.
If “NORMAL” is not displayed when 1 or less indicator is display, refer to page 11-2, “Error in
measured value”.

e

e

e

e

e “Battery alarm”

When activating this instrument on the built-in battery, check if the BATTERY ALARM (

) is not

displayed. If “BATTERY ALARM” is displayed, the power is turned OFF in about 20 minutes.
When charging the battery, refer to page 4-1 “To charge the battery”.

r

r

r

r

r Unit of integration

When changing the unit of integration, refer to page 9-6, “TOTALIZE: when performing the
integration process of measured data (totalize)”.

t

t

t

t

t Unit of indication value

To change the units of flow rate and flow
velocity on the MEASURE screen;
q

Move the cursor to “UNIT” by pressing the

or

key.

w

Press the

ENT

key, and the screen

appears, enabling the unit of flow rate to be

selected. Select any unit by pressing the

or

key and press the

ENT

key.

y

y

y

y

y Status display of integration

Meaning of display

STOP:

Not totalized

RUN:

Totalizing in progress

To start the action of integration, refer to page
9-6, “TOTALIZE”.

u

u

u

u

u “Clock set”

This instrument has a timer function. For the
timer to set the time, refer to page 9-15,
“CLOCK SET”.
The timer function should be used based on this
watch.

i

i

i

i

i Reset

The integration value can be set to 0 or “any

other numeric value”. To reset the integration

value, point the cursor to “RESET” by pressing

the

or

key, and then press the

ENT

key.

When you want to reset to any value or 1000 for

example, refer to page 9-7, “To set reset data”.

x10

x10

RESET

STOP

STOP

RESET

RESET

FLOW RATE

UNIT:

m3/h

VELOCITY

UNIT:

m/s

+ TOTAL

UNIT:

ml

– TOTAL

UNIT:

ml

MEASURE

NORMAL

x10

x10

RESET

STOP

STOP

RESET

RESET

FLOW RATE

UNIT:

m3/h

VELOCITY

UNIT:

m/s

+ TOTAL

UNIT:

ml

– TOTAL

UNIT:

ml

MEASURE

NORMAL

L / m i n

L / h

M L / d

m 3 / s

m 3 / m i n

M m 3 / d

B B L / s

B B L / m i n

B B L / h

MBBL/d
m/s
AI %
AO %

UNIT

UNIT

m 3 / h

L / s

Point the cursor to
UNIT of flow rate
and press the
key, and Menu
screen is displayed.

Point the cursor to
RESET and press
the key,
and totalize is
reset.

ENT

ENT

BULLETIN F-68