q “Indicator”
Shows the intensity of ultrasonic receiving signal.
Check if 2 or more indicators are displayed.
If one or less indicator is displayed, raise the transmission voltage level as shown on page 6-12.
When the sensor is not connected, the indicator may be lighted by sensing noises. But, this is not
an error.
w
w
w
w
w “Status display”
Check if “NORMAL” is displayed. If the sensor is not connected, other messages may be displayed.
This is not an error.
In case other message is displayed after installing and connecting the sensor, take corrective actions
according to page 9-33, “System check function”.
If “NORMAL” is not displayed when 1 or less indicator is display, refer to page 11-2, “Error in
measured value”.
e
e
e
e
e “Battery alarm”
When activating this instrument on the built-in battery, check if the BATTERY ALARM (
) is not
displayed. If “BATTERY ALARM” is displayed, the power is turned OFF in about 20 minutes.
When charging the battery, refer to page 4-1 “To charge the battery”.
r
r
r
r
r Unit of integration
When changing the unit of integration, refer to page 9-6, “TOTALIZE: when performing the
integration process of measured data (totalize)”.
t
t
t
t
t Unit of indication value
To change the units of flow rate and flow
velocity on the MEASURE screen;
q
Move the cursor to “UNIT” by pressing the
or
key.
w
Press the
ENT
key, and the screen
appears, enabling the unit of flow rate to be
selected. Select any unit by pressing the
or
key and press the
ENT
key.
y
y
y
y
y Status display of integration
Meaning of display
STOP:
Not totalized
RUN:
Totalizing in progress
To start the action of integration, refer to page
9-6, “TOTALIZE”.
u
u
u
u
u “Clock set”
This instrument has a timer function. For the
timer to set the time, refer to page 9-15,
“CLOCK SET”.
The timer function should be used based on this
watch.
i
i
i
i
i Reset
The integration value can be set to 0 or “any
other numeric value”. To reset the integration
value, point the cursor to “RESET” by pressing
the
or
key, and then press the
ENT
key.
When you want to reset to any value or 1000 for
example, refer to page 9-7, “To set reset data”.
x10
x10
RESET
STOP
STOP
RESET
RESET
FLOW RATE
UNIT:
m3/h
VELOCITY
UNIT:
m/s
+ TOTAL
UNIT:
ml
– TOTAL
UNIT:
ml
MEASURE
NORMAL
x10
x10
RESET
STOP
STOP
RESET
RESET
FLOW RATE
UNIT:
m3/h
VELOCITY
UNIT:
m/s
+ TOTAL
UNIT:
ml
– TOTAL
UNIT:
ml
MEASURE
NORMAL
L / m i n
L / h
M L / d
m 3 / s
m 3 / m i n
M m 3 / d
B B L / s
B B L / m i n
B B L / h
MBBL/d
m/s
AI %
AO %
UNIT
UNIT
m 3 / h
L / s
Point the cursor to
UNIT of flow rate
and press the
key, and Menu
screen is displayed.
Point the cursor to
RESET and press
the key,
and totalize is
reset.
ENT
ENT
BULLETIN F-68