beautypg.com

3 error in measured value – Dwyer PUX User Manual

Page 82

background image

11- 2

State

Cause

Indication of measured
value is negative (-).

Measured value
fluctuated widely
though flow rate is
constant.

Measured value
remains the same
though flow rate is
changing.

Connect correctly.

Shift measurement location to
the site where 10D and 5D can
be secured on the upstream and
downstream sides.

Mount the instrument with a
clearance of 30D or more.

Extend response time through
damping setting.

11.3 Error in measured value

Remedy

• Connection between the main unit

and sensors (upstream sensor and
downstream sensor) is reverse.

• Fluid is actually flowing in the (-)

direction.

• Straight pipe portion is inadequate.

• A flow disturbing factor such as

pump or valve is provided in the
vicinity.

• Pulsation is occurring actually.

• Measured value is held because ultrasonic

wave cannot be propagated into a pipe.

1. Incomplete installation

• Piping specifications are wrong.
• Sensor is mounted at the welded

part.

• Sensor mounting dimension is

wrong.

• Grease application at sensor

mounting is incomplete.

• Sensor connector is not

connected completely.

• Pipe surface is contaminated.

2. Problem on pipe or fluid

• When V method is used, it should

be changed to Z method.

• When sensor extension cable is

used, it should be avoided.

• If error persists, check and

eliminate the true cause as
instructed below.

Fluid is not filled.

After check, separate the sensor
once. Apply the grease again
and remount the sensor with a
slight shift.

Find the location on the same
pipe line, where fluid is filled
up and move the sensors to that
location.
Mount the sensors at the lowest
location on pipe line.

BULLETIN F-68