9 - 6

9.1.6 TOTALIZE: when performing the integration process of

measured data (totalize)

[Operation]

Select "TOTALIZE" by the

or

key and press

the

ENT

key. The "TOTALIZE" screen will appear.

q

q

q

q

q

To start integration

To open the Integration Output Selection screen, move the
cursor to “MODE” and press the

ENT

key.

To set the integration output, select the integration output and
press the

ENT

key.

Setting is required here when using integration output.

Mode

Description

OFF

Manual

Quick timer

Integration stop

Instant integration start

Timer

Even if the timer settig is performed after scheduled start

time already passed, integration will not be carried out.

Integration starts after the time of setting, integration is

performed within the time selected from the menu, and it

stops automatically after the time passed.

Hour: minute

ENT

ENT

Set the time of integration to start and stop. Point the

cursor to "TIMER START", and then press the key.

Integration starts and stops automatically.

Point the cursor to "TIMER START", and integration will

not be started untill the key is pressed.

TIMER

END DATE / TIME

START DATE / TIME

TIMER START

01-01 00 : 00

01-01 00 : 00

QUICK

01 :

: 00

00 : 30

00 : 30

02 :

: 00

01 :

: 30

03 :

: 00

02 :

: 30

SITE SETUP

MODE

TOTALIZE

MANUAL

OFF

OFF

TOTALIZE

SITE SETUP

RESET DATA

MODE

MODE

OFF

0

UNIT

m l

QUICK TIMER

00 : 30

TIMER

ESC

SITE SETUP

SITE SETUP

PIPE PARAMETER

ZERO ADJUST

RESPONSE SET

CALIBRATION

CUT OFF

TOTALIZE

0 s e c

0 . 0 0 0 m/s

--- SENSOR SPACING ---

0.0mm

PARAMETER MAMORY

MANUAL ZERO

BULLETIN F-68