Bulletin f-68 – Dwyer PUX User Manual

Page 78

background image

9 - 37

(2) Check to judge whether ultrasonic receiving signal waveform is normal or

not

q

Normal waveform

The receiving waveform above is free of noise,
and so normal measurement is allowed.

w

Abnormal waveform

The above receiving waveform is not covered
within the range of the ultrasound waveform. It
is displayed as “CALCULATION ERROR” or
“RECEIVED SIGNAL ERROR”.

Check the pipe setting and sensor mounting
dimensions.

When ultrasonic receiving signal waveform is
weakened by the effect of rust in the pipe,
abnormal waveform may result. Raise the
transmission voltage and perform measurement.
(See page 6-12).

SYSTEM CHECK

DATA / 1DIV

CURSOR

UP STREAM WAVE

2 0

5 0 0 0

272

SOURCE

1 0 4 4 6

SOUND SPEED

SOUND SPEED

1 5 2 1 m / s

1 5 2 1 m / s

Peak value of about
15000 to 20000

TIME / 1DIV

TIME / 1DIV

5000

SYSTEM CHECK

DATA / 1DIV

UP STREAM WAVE

2 0

5 0 0 0

112

SOURCE

- 6 2 8 9

SOUND SPEED

SOUND SPEED

1 6 6 9 m / s

1 6 6 9 m / s

CURSOR

TIME / 1DIV

TIME / 1DIV

Receiving wave-
form is cut off.

Receiving
range

BULLETIN F-68