Bulletin f-68 – Dwyer PUX User Manual
Page 60
9 - 19
(2) INPUT RANGE: when setting the range of input current
q
Press the
or
key on the “ANALOG IN/OUT” page and select “INPUT RANGE” from
“INPUT”.
Press the
ENT
key, and you are ready to adjust input range.
w
Move the digit by the
or
key and enter numeric values by pressing the
or
key. After
entry, press the
ENT
key to set the range.
(3) DISPLAY: when displaying values by entering analog signals (4 to 20 mA)
of other devices
q
Press the
or
key on the “ANALOG IN/OUT” page and select “DISPLAY” from “INPUT”.
Press the
ENT
key to display present instantaneous flow and input value. Selection of “INPUT”
indicates a displayed value equal the input current converted into input (%)
×
OUTPUT RANGE.
ANALOG IN / OUT
ANALOG IN / OUT
INPUT
ANALOG
0 . 0 0 0 E 0
ADJUST
OUTPUT
DISPLAY
RANGE UNIT
OUTPUT RANGE
OUTPUT MODE
BURN-OUT
ADJUST
m/s
HOLD
1 . 0 0 0 E 0
0 . 8 - 4 - 2 0 m A
DISPLAY
INPUT RANGE
Index is settable in
range from –9 to +9
Input:
4 mA to 20 mA
Display: 0 to set value
[Example]
To display inputs of 4 to
20 mA within the range
of 0 to 100;
0.000 E0
Mantissa
Index
1.000 E2
1.000 10
2
×
ANALOG IN / OUT
ANALOG IN / OUT
INPUT
ANALOG
0 . 0 0 0 E 0
ADJUST
OUTPUT
DISPLAY
RANGE UNIT
OUTPUT RANGE
OUTPUT MODE
BURN-OUT
ADJUST
m/s
HOLD
1 . 0 0 0 E 0
0 . 8 - 4 - 2 0 m A
INPUT RANGE
DISPLAY
ANALOG IN / OUT
DISPLAY
INPUT
8.000E0
ESC
Display = INPUT (%)
×
OUTPUT RANGE/100
Input current
-
4mA
16mA
×
100
BULLETIN F-68