(2) INPUT RANGE: when setting the range of input current

q

Press the

or

key on the “ANALOG IN/OUT” page and select “INPUT RANGE” from

“INPUT”.

Press the

ENT

key, and you are ready to adjust input range.

w

Move the digit by the

or

key and enter numeric values by pressing the

or

key. After

entry, press the

ENT

key to set the range.

(3) DISPLAY: when displaying values by entering analog signals (4 to 20 mA)

of other devices

q

Press the

or

key on the “ANALOG IN/OUT” page and select “DISPLAY” from “INPUT”.

Press the

ENT

key to display present instantaneous flow and input value. Selection of “INPUT”

indicates a displayed value equal the input current converted into input (%)

×

OUTPUT RANGE.

ANALOG IN / OUT

ANALOG IN / OUT

INPUT

ANALOG

0 . 0 0 0 E 0

ADJUST

OUTPUT

DISPLAY

RANGE UNIT

OUTPUT RANGE

OUTPUT MODE

BURN-OUT

ADJUST

m/s

HOLD

1 . 0 0 0 E 0

0 . 8 - 4 - 2 0 m A

DISPLAY

INPUT RANGE

Index is settable in
range from –9 to +9
Input:

4 mA to 20 mA

Display: 0 to set value

[Example]
To display inputs of 4 to
20 mA within the range
of 0 to 100;

0.000 E0

Mantissa

Index

1.000 E2

1.000 10

2

×

ANALOG IN / OUT

ANALOG IN / OUT

INPUT

ANALOG

0 . 0 0 0 E 0

ADJUST

OUTPUT

DISPLAY

RANGE UNIT

OUTPUT RANGE

OUTPUT MODE

BURN-OUT

ADJUST

m/s

HOLD

1 . 0 0 0 E 0

0 . 8 - 4 - 2 0 m A

INPUT RANGE

DISPLAY

ANALOG IN / OUT

DISPLAY

INPUT

8.000E0

ESC

Display = INPUT (%)

×

OUTPUT RANGE/100

Input current

-

4mA

16mA

×

100

BULLETIN F-68