Detector (type: fld), Functions, Bulletin f-68 – Dwyer PUX User Manual
Page 100
15 - 9
•
Enclosure case:
Plastic case (color: gray)
•
Dimensions:
H240 x W127 x D70mm (without printer)
H359 x W127 x D70mm (with printer)
•
Mass:
1.5kg (without printer)
2.0kg (with printer)
Detector (Type: FLD)
•
Mounting method:
Mounting on outside of already constructed pipe
•
Sensor mounting method: V or Z method
•
Mounting belt /wire:
Small diameter sensor, plastic cloth belt
Small sensor, plastic cloth belt
Large sensor, stainless wire
High-temperature sensor, stainless belt
•
Acoustic coupler:
Silicone grease
•
Signal cable:
Special type coaxial cable
•
Connection:
Converter; BNC connector
Small diameter sensor; BNC connector
Small sensor; BNC connector
Large sensor; terminal screws
High-temperature sensor; BNC connector
•
Ambient temperature: –20 to +60
°
C
•
Ambient humidity:
Large sensor, 100% RH or less
Other, 90% RH or less
•
Type of enclosure:
Large sensor: immersion-proof type (IP67 or equivalent)
Other:
drip-proof type (IP52 or equivalent)
•
Material:
•
Dimensions/mass:
Functions
•
Display language:
English
/
Japanese (Katakana)
/German/French, selectable
•
Instantaneous value display function:
Two of velocity, flow rate (with flow direction) and analog input, simultaneous
display
Unit: Velocity m/s
Flow rate L/s, L/min, L/h, ML/d, m
3
/s, m
3
/min, m
3
/h, Mm
3
/d
•
Total value display function: Forward and reverse total values, simultaneous display
Unit:
mL, L, m
3
, km
3
, Mm
3
•
Clock display function: Time (year, month, day, hour, minute) display and setting
•
Damping:
0 to 99s (time constant)
•
Low flow cut:
0 to 1.000m/s
•
Output setting function: Current output scaling, output type, burnout setting and calibration
•
Communication function: Velocity, flow rate, totals, analog input, status, logging data transmission on
request
•
Memory function:
Site data (place, piping, fluid, sensor mounting method, type of sensor) up to 20
places and a maximum of 40000 data (time, velocity, flow rate, totals, analog
input, status) can be stored in memory.
•
Waveform display function: Display of bi-directional received waveforms
•
Graph display function: Display of velocity, flow rate trend graph
•
Printing function:
Printout of screen, fixed cycle printout (time, velocity, flow rate, totals, analog
input, status), logging data, trend graph, and waveforms by using integral printer
(option)
Kind
Sensor case
Mounting bracket
Small diameter
Plastic
Aluminum alloy + Plastic
Small type
Plastic
Aluminum alloy + Plastic
Large type
Plastic
—
High temperature
304SS
Aluminum alloy + 304SS
Kind
Dimensions (HxWxD)
Small diameter
420 x 53 x 90mm
0.6kg
Small type
540 x 53 x 90mm
0.8kg
Large type
104 x 93 x 62mm
1.4kg
High temperature
530 x 52 x 205mm
1.7kg
Mass
(Note)
Note) mass of both sensors
BULLETIN F-68