7 - 3

7.2

Selection of detector

(1) Selection of mounting methods

There are 2 methods for mounting the detector; V method and Z method. For the mounting space, see

the following sketch.

Employ the Z method in the following cases.
• Mounting space need be saved (mounting space of the Z method is about one half of the V

method's).

• Turbid fluid such as sewage is to be measured.
• Pipe has mortar lining.
• A thick film of scale may have been formed on the inner surface of pipe because it is old.

(2) Detector selection standards

The Z method for large size sensor is recommended for outer diameter 300mm or more.
PSX2 D should be used as much as possible for pipes such as old pipes, cast iron pipes, and mortar
lining pipes, through which it is difficult for ultrasonic signals to pass.

Length of frame of each sensor

Frame

Length of frame of each sensor

L

L

L: Mounting dimension

mounting dimension for
sensor displayed on the
SITE SETUP screen

D

D

Detector

L=About D

V method

Detector

D

L=About D/2

Z method

V method

Z method

Mounting method

Type

Diameter

Temperature

PSX2 B

13

100mm (V method)

-40

+100

°

C

-40

+100

°

C

-40

+200

°

C

-40

+80

°

C

Detector

PSX2 A/E

50

350mm (V method)

300

400mm (Z method)

PSX2 HT

50

350mm (V method)

PSX2 D

200

3000mm (V method)

200

6000mm
(Z method)

300

400mm (Z method)