Bulletin f-68 – Dwyer PUX User Manual
Page 19
6 - 2
r
Outline of PIPE PARAMETER
t
Display of mounting dimensions
When the settings are completed on the “PIPE PARAMETER” screen, press the
ESC
key to display the
“SITE SETUP” menu screen.
At the last line the “SENSOR SPACING” value is displayed.
Install the sensor according to chapter 7. MOUNTING OF DETECTOR and the mounting
dimension is as displayed on the last line.
SITE SETUP
SITE SETUP
96-06-24 10:57
SITE SETUP
OUTER DIAMETER
PIPE MATERIAL
WALL THICKNESS
LINING MATERIAL
KIND OF FLUID
SENSOR MOUNTING
CARBON STEEL
0.01 m m
PIPE PARAMETER
PIPE PARAMETER
13.00 m m
NO LINING
V
PSX2 B
SENSOR TYPE
LINING THICKNESS
0.01mm
WATER
SITE NAME
SITE NAME
X
1
TRANS, VOLTAGE
Enter site name (name of pipe to be used for measurement)
→
Page 6 - 3
Set external dimensions of pipe
→
Page 6 - 4
Set pipe material
→
Page 6 - 5
Set pipe thickness
→
Page 6 - 6
Set lining material
→
Page 6 - 7
Set lining thickness
→
Page 6 - 8
Set kind of fluid
→
Page 6 - 9
Set selection of sensor mounting method
→
Page 6 - 10
Set type of sensor
→
Page 6 - 11
Set transmission voltage
→
Page 6 - 12
SITE SETUP
SITE SETUP
96-06-24 10:57
SITE SETUP
OUTER DIAMETER
PIPE MATERIAL
WALL THICKNESS
LINING MATERIAL
KIND OF FLUID
SENSOR MOUNTING
CARBON STEEL
0.01 m m
PIPE PARAMETER
PIPE PARAMETER
13.00 m m
NO LINING
V
PSX2 B
SENSOR TYPE
LINING THICKNESS
0.01mm
WATER
SITE NAME
X
1
TRANS, VOLTAGE
TRANS, VOLTAGE
95-06-27 14:49
SITE SETUP
SITE SETUP
SITE SETUP
PARAMETER MEMORY
PIPE PARAMETER
ZERO ADJUST
RESPONSE SET
CALIBRATION
CUT OFF
TOTALIZE
MANUAL ZERO
0 s e c
0 . 0 0 0 m / s
--- SENSOR SPACING ---
125.7mm
ESC
Mounting
dimension
BULLETIN F-68