5 cut off: output cut off at low flow rate, Bulletin f-68 – Dwyer PUX User Manual

Page 46

background image

9 - 5

9.1.5 CUT OFF: output cut off at low flow rate

(low flow cutoff function)

When flow rate is extremely low, its output can be
cut off. (range: 0 to 1.000 m/s)
If fluid in the pipe is moving due to convection, etc.,
even though the valve is closed, this flowmeter
outputs a measured value. Therefore, values below
an appropriate level should be cut off.

0.010m/s

Flow velocity

- 0.010m/s

Output

[Operation]

q

Press the

or

key on the SITE SETUP page

and select CUT OFF. Then, press the

ENT

key,

and the cursor moves to the set item. Output cut
off point is settable.

w

Move the digit by pressing the

or

key and

enter a numeric value by pressing the

or

key.

After entry, press the

ENT

key.

SITE SETUP

SITE SETUP

SITE SETUP

PIPE PARAMETER

ZERO ADJUST

RESPONSE SET

CALIBRATION

TOTALIZE

MANUAL ZERO

0 s e c

--- SENSOR SPACING ---

0.0mm

PARAMETER MEMORY

CUT OFF

0.0 0 0 m/ s

BULLETIN F-68