6.11 Transmission voltage (used when an indicator is 1 or

less during measurement)

Press the

key to invert the “TRANS. VOLTAGE”.

Press the

ENT

key. The screen is ready to allow the

selection of the transmission voltage level.

Use the

or

key to select the level.

Select “4 TIMES” or “8 TIMES” generally.

If the indicator cannot be set to 2 or more
with the level at “8 TIMES”, ultrasonic wave
may be attenuated due to contamination or
scales deposited on the piping external and
internal surfaces. Change measurement
location.

T h e i n d i c a t o r w i l l b e u p d a t e d o n t h e
“MEASURE” screen only.

If less than 2 indicators (intensity of receiving

waveform) are displayed on the “MEASURE” screen,

raise the transmission voltage.

CAUTION

ENT

ENT

SITE SETUP

SITE SETUP

SITE SETUP

PIPE PARAMETER

PIPE PARAMETER

V

SITE NAME

OUTER DIAMETER

PIPE MATERIAL

WALL THICKNESS

LINING MATERIAL

KIND OF FLUID

SENSOR MOUNTING

PSX2 A

SENSOR TYPE

LINING THICKNESS

4 TIMES

TRANS. VOLTAGE

TRANS. VOLTAGE

SITE SETUP

SITE SETUP

SITE SETUP

PIPE PARAMETER

PIPE PARAMETER

SITE NAME

OUTER DIAMETER

PIPE MATERIAL

PIPE MATERIAL

WALL THICKNESS

LINING MATERIAL

KIND OF FLUID

SENSOR MOUNTING

SENSOR TYPE

LINING THICKNESS

TRANS, VOLTAGE

WATER

1.25 m m

MORTAR

1.25mm

CAST IRON

318.50 m m

PIPE NO.1

TRANS. VOLTAGE

TRANS. VOLTAGE

2 TIMES

1 TIME

4 TIMES

4 TIMES

8 TIMES

95-06-27 11:49

x10

x10

RESET

STOP

FLOW RATE

UNIT:

m3/h

VELOCITY

UNIT:

m/s

+ TOTAL

UNIT:

ml

MEASURE

Indicator

Example) When transmission voltage

is set to eight-fold

8

TIMES

ENT

8

TIMES

TRANS. VOLTAGE

TRANS. VOLTAGE

2

TIMES

4

TIMES

1

TIME

2

TIMES

4

TIMES

1

TIME