6.5.4 Density calibration (menu C1.2.1)

Mass flowmeters are calibrated for density in the factory. Density calibration is based on 2
calibration points. In the factory, air and water are used under reference conditions. The result
of this calibration is saved in the converter electronics and stored in the factory settings.
Regardless, various applications require maximum accuracy which can only be achieved with
on-site calibration.

Available options:

Example of 1 point calibration with town water

Option

Explanation

1 Point Calibration

One of the 2 saved calibration points is replaced by customer calibration.

The signal converter decides which of the 2 calibration points is modified.

2 Point Calibration

The user recalibrates both calibration points.

Default

The signal converter reactivates the factory settings for the density

calibration.

Manual

The user can read the current density calibration points and enter them if

necessary (according to sensor calibration certificate).

Key

Display

Description and setting

>

A

Quick Setup

Press and hold for 2.5 s, then release

the key.

2 x ↓

C

Setup

2 x >

C1.1

Calibration

C1.2

Density

2 x >

C1.2.1

Density Calib.

Break

Press ^ to exit density calibration

menu.

C1.2.1

Density Calibration?

Default

Press ^ for density calibration.

Factory calibration on OK (6 x ^).

C1.2.1

Density Calibration?

Manual

Press ^ to read or change the

current density calibration value.

C1.2.1

Density Calibration?

2 Point Calibration

Press ^ to start 2 point calibration.

C1.2.1

Density Calibration?

1 Point Calibration

Press ^ to start 1 point calibration.

^

C1.2.1

DCF1

XXXXXXXXX

Press ↓ until you reach town water.

Press ↓

until

C1.2.1

DCF1

Town Water

Press ^ to start town water

calibration.

^

C1.2.1

Single Pt. Calib.

Break

C1.2.1

Single Pt. Calib.

OK

Press ^ to start single point

calibration.

^

C1.2.1

Perform Calibration

Passed

5 x ^

Save Configuration?

Yes

^

Measuring Mode