Operation – KROHNE MFC 300 EN User Manual

Page 80

background image

6

OPERATION

80

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

A4 Station Address

A4

Station Address

For Profibus / FF / Modbus devices.

A5 Digital Outputs

A5

Digital Outputs

Valid for all pulse outputs (terminals A, B and/or D) and totaliser 1.

A5.1

Measurement

1) Select measurement: volume flow / mass flow / concentration flow 1

2) Use for all outputs? (also use this setting for Fct. A5.2...A5.5!)

Setting: no (only for pulse output D) / yes (for all digital outputs)

A5.2

Pulse Value Unit

Selection of the unit from a list, depending on the measurement.

A5.3

Value p. Pulse

1) Setting for pulse output D (volume or mass value per pulse)

Setting: xxx.xxx in l/s or kg/s

2) Use for all outputs? Make setting, see Fct. A5.1 above!

A5.4

Low Flow Cutoff

1) Setting for pulse output D (sets output value to "0")

Setting: x.xxx ± x.xxx% (range: 0.0…20%)

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

2) Use for all outputs? Make setting, see Fct. A5.1 above!

A6 GDC IR interface

A6

GDC IR interface

After this function has been activated an optical GDC adapter can be

connected to the LC display. If approximately 60 seconds pass without a

connection being established or after the adapter is removed, then the

function is exited and the optical keys are active once again.

break (exit function without connection)

activate (the IR interface (adapter) and interrupt the optical keys)

A7 Zero Calibration

A7

Zero Calibration

Zero calibration, calibration sequence as described in Fct. C1.1.1...1.1.4

A8 Operation Mode

A8

Operation Mode

Set the operating mode.

Setting: measure / stop / standby

For detailed information refer to

Mode (menu A8) on page 98.

No.

Function

Settings / descriptions

.book Page 80 Monday, February 6, 2012 11:01 AM