SERVICE

7

127

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

7.2.3 OPTIMASS 3000

The specified values are for guidance only.
Damaged magnetic coil: inductance values in parentheses.

• Driver = black/violet and grey/orange
• Sensor A = white and yellow
• Sensor B = green and yellow
• Pt500 = red and blue (530…550 Ω) at ambient temperature
• Resistance values outside of the ranges specified above may indicate a circuit error. The

device may be in "start-up" mode or indicate measuring errors.

• All circuits should be isolated with respect to earth (device housing) and each other >20 MΩ.
• An exciter/sensor coil short circuit may cause the meter to be in "start-up" mode.

Size

Inductance (mH)

Resistance (Ohm)

Driver

Sensor A/B

Driver

Sensor A/B

01

1.2 (0.6)

7.4

57

107

03 / 04

2.4 (1.2)

10.1

47

135

INFORMATION!
Failure of two or more of the upper circuits may indicate a measuring tube error. There may be
product in the housing. If this is the case, depressurise

depressurise

depressurise

depressurise the process line and immediately remove

the measuring device from the process line.