6.7 Function tests and troubleshooting
Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)
Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)
Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)
Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)
Stores the minimum and maximum value for the temperature occurring during sensor operating
time.
Application problems that may indicate errors in the signal converter:
• An improperly closed shut-off valve during zero calibration results in high calibration values
• Air/gas bubbles result in high energy levels and high calibration values
• Product deposits on the inside of the measuring tube lead to higher/lower density indications
and high calibration values
Common errors (with corresponding symptoms):
• Slightly eroded or corroded measuring tube
- Erroneous density measurement
- High frequency
- Measurement error at low mass flow
• Eroded or corroded measuring tube (product in the housing)
- Measuring tube does not start
- Low resistance to earth with conducting product
• Open circuits from activation and sensor coils, resistance thermometers (RTD) or strain
gauge
- Can be measured with ohmmeter
Sensor type
Operating temperature
Minimum
Maximum
OPTIMASS 1000
-40°C / -40°F
130°C / 266°F
OPTIMASS 2000
OPTIMASS 3000
(Stainless steel or Hastelloy
®
)
-30°C / -22°F
150°C / 302°F
OPTIMASS 7000 (titanium)
-40°C / -40°F
150°C / 302°F
OPTIMASS 7000
(Hastelloy
®
/tantalum)
0°C / 32°F
100°C / 212°F
OPTIMASS 7000 (stainless steel)
0°C / 32°F
100°C / 212°F
130°C / 266°F
OPTIMASS 8000
(depends on version)
-195°C / -310°F
230°C / 446°F