6.7 Function tests and troubleshooting

Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)

Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)

Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)

Min. and max. recorded temperature (menu C1.5.1 / C1.5.2)

Stores the minimum and maximum value for the temperature occurring during sensor operating
time.

Application problems that may indicate errors in the signal converter:

• An improperly closed shut-off valve during zero calibration results in high calibration values
• Air/gas bubbles result in high energy levels and high calibration values
• Product deposits on the inside of the measuring tube lead to higher/lower density indications

and high calibration values

Common errors (with corresponding symptoms):

• Slightly eroded or corroded measuring tube

- Erroneous density measurement
- High frequency
- Measurement error at low mass flow

• Eroded or corroded measuring tube (product in the housing)

- Measuring tube does not start
- Low resistance to earth with conducting product

• Open circuits from activation and sensor coils, resistance thermometers (RTD) or strain

gauge
- Can be measured with ohmmeter

Sensor type

Operating temperature

Minimum

Maximum

OPTIMASS 1000

-40°C / -40°F

130°C / 266°F

OPTIMASS 2000

OPTIMASS 3000
(Stainless steel or Hastelloy

®

)

-30°C / -22°F

150°C / 302°F

OPTIMASS 7000 (titanium)

-40°C / -40°F

150°C / 302°F

OPTIMASS 7000
(Hastelloy

®

/tantalum)

0°C / 32°F

100°C / 212°F

OPTIMASS 7000 (stainless steel)

0°C / 32°F

100°C / 212°F

130°C / 266°F

OPTIMASS 8000

(depends on version)

-195°C / -310°F

230°C / 446°F