Operation – KROHNE MFC 300 EN User Manual

Page 116

background image

6

OPERATION

116

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

Simulation of the measured values

S 2 Phase Flow

2 phase signal is above the programmed

threshold.

Check the process conditions (air).

S Interface PCB Fault

Error detected during self-monitoring of

interface card. Possible causes:

temperature in converter housing too high

or blown fuse.

Check whether the converter is exposed to

direct sunlight. Check the temperature in

Fct. B2.15. Replace electronic unit.

Messages on the display Description

Actions

Status: C _ _ _ _ _

Output values partially simulated or
fixed

Maintenance required.

C Checks In Progress

Test mode of the device. Measured values

are possibly simulated values or values with

fixed settings.

Message depending on the situation via
HART

®

or FDT. Depiction via display if

outputs are held by control input or set to

zero.

C Test XXXXX

Test of relevant unit activated.

C Standby Mode

Device is in standby mode.

Check control input settings in A8.

C Sensor Electronics

A test function in the sensor electronics (SE)

has been activated.

C Zero Calibration

Zero calibration performed. Flow

measurement interrupted.

Messages on the display Description

Actions

Status: S _ _ _ _ _

Out of specification, measurement
continues, accuracy possibly less.

Maintenance required.

.book Page 116 Monday, February 6, 2012 11:01 AM