Operation – KROHNE MFC 300 EN User Manual

Page 114

background image

6

OPERATION

114

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

Application Error

F SE Data Different

Sensor electronics (SE) and converter

electronics (BE) have different parameters.

The electronic insert has probably been

replaced.

Copy the parameters from SE to BE or vice

versa in Fct. C1.6.3. For detailed information

refer to

Replacing the sensor or converter

electronics on page 122.

F SE Drive Failure

Error in sensor electronics (SE), drive

amplitude no longer controllable.

Replace electronic unit.

F SE Wiring Error

Fault in wiring (remote version)

Check wiring and rectify.

F Interface PCB Failure

RAM or ROM error detected.

Replace electronic unit.

F Hardware Settings (also

when changing modules)

The set hardware parameters do not match

the identified hardware. A dialogue appears

in the display.

Answer queries in dialogue mode, follow

directions.

After module change, confirm query for

changed configuration. If device

configuration unchanged: defective, replace

electronic unit.

F Hardware Detection

Existing hardware cannot be identified.

Defective or unknown modules.

Replace electronic unit.

F RAM/ROM error IO1

A RAM or ROM error is detected during the

CRC check.

Defective, replace electronic unit or

input/output module (I/O module).

F RAM/ROM error IO2

F Fieldbus

Malfunction of the Fieldbus interface.

Defective, replace electronic unit or

input/output module (I/O module).

Messages on the display Description

Actions

Status: F _ _ _ _ _

Application fault, device OK, but
measured values affected.

Application test or operator action
necessary.

F Application Error

Application-dependent fault, but device is

OK.

Group message, when errors as described

below or other application errors occur.

F Sensor Exceeding Limit

Mass flow is greater than max. flowrate.

Accuracy is not guaranteed!

Reduce flowrate or increase meter size.

F Open Circuit A

Load on current output A/B/C too high,

effective current too low.

Current not correct, mA output cable has

open circuit or load too high. Check cable,

reduce load (set < 1000 ohm).

F Open Circuit B

F Open Circuit C

F Over Range A

The current or the corresponding measured

value is limited by a filter setting.

Check with Fct. C3.1 hardware or sticker in

terminal compartment, which output is

connected to the terminal.

If current output: extend Fct. C3.x.6 range

and Fct. C3.x.8 limitation.

If frequency output: extend values in

Fct. C3.x.5 and Fct. C3.x.7.

F Over Range B

F Over Range C

F Over Range D

F Over Range A

The pulse rate or the corresponding

measured value is limited by a filter setting.

Or the demanded pulse rate is too high.

F Over Range B

F Over Range C

F Over Range D

F Wiring A

Wiring error.

Check connections at terminals A or B.

F Wiring B

F Stop Mode

Device is in stop mode.

Check Fct A8.

Messages on the display Description

Actions

Status: F _ _ _ _ _

Operational fault in device, mA output ≤
3.6 mA or set fault current (depending
on the seriousness of the fault), status
output open, pulse / frequency output:
no pulses

Repair necessary.