OPERATION

6

117

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

Information

Messages on the display Description

Actions

Status: I _ _ _ _ _

Information (current measurement OK)

I Totaliser 1 Stopped

This is totaliser 1 or FB2 (with Profibus). The

totaliser has stopped.

If totaliser to continue counting, activate

"yes" in Fct. C4.y.9 (start totaliser).

I Totaliser 2 Stopped

This is totaliser 2 or FB3 (with Profibus). The

totaliser has stopped.

I Totaliser 3 Stopped

This is totaliser 3 or FB4 (with Profibus). The

totaliser has stopped.

I Power Fail

The device was not in operation for an

unknown period of time, because the power

was switched off. This message is for

information only.

Temporary power failure. Totaliser did not

run during it.

I Control Input A act.

This message appears when the control

input is active. This message is for

information only.

I Control Input B act.

I Over Range Display 1

1st line on page 1 (2) of display limited by

filter setting.

Menu display Fct. C6.3 and/or C6.4, select

1st or 2nd meas. page and increase values

in functions C6.z.3 range and/or C6.z.4

limitation

I Backplane Sensor

The data on the backplane are not usable

because they have been generated with an

incompatible version.

I Backplane Settings

The global settings on the backplane are not

usable because they have been generated

with an incompatible version.

I Backplane Difference

The data on the backplane differ from the

data in the display. If the data are usable, a

dialogue is indicated in the display.

I Optical Interface

The optical interface is being used. The keys

on the local display are not in operation.

The keys are ready for operation again

approx. 60 sec. after the end of the data

transfer/removal of the optical interface.

I Write Cycles Overfl.

The maximum number of write cycles of the

EEPROM or FRAMS on the Profibus DP PCB

has been exceeded.

I Baudrate Search

The baudrate of the Profibus DP interface is

searched for.

I No Data Exchange

There is no data exchange between the

signal converter and the Profibus.

.book Page 117 Monday, February 6, 2012 11:01 AM