beautypg.com

Service – KROHNE MFC 300 EN User Manual

Page 128

background image

7

SERVICE

128

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

7.2.4 OPTIMASS 7000

The specified values are for guidance only.

• Driver = black and grey
• Sensor A = white and yellow
• Sensor B = green and violet
• Pt500 = red and blue (530…550 Ω) at ambient temperature
• Resistance values outside of the ranges specified above may indicate a circuit error. The

device may be in "start-up" mode or indicate measuring errors.

• All circuits should be isolated with respect to earth (device housing) and each other >20 MΩ.
• An exciter/sensor coil short circuit may cause the meter to be in "start-up" mode.

Size

Resistance (Ohm)

Driver

Sensor A/B

06 / 10

40

142

15

49

142

25

41

142

40 / 50 / 80

100

142

Strain MT = red and brown

OPTIMASS 7000 - all sizes

420...560 Ω at ambient

temperature

Strain IC = brown and orange

OPTIMASS 7000 - 06...10

225...275 Ω at ambient

temperature

OPTIMASS 7000 - 15...80

Short circuited

INFORMATION!
Failure of two or more of the upper circuits may indicate a measuring tube error. There may be
product in the housing. If this is the case, depressurise

depressurise

depressurise

depressurise the process line and immediately remove

the measuring device from the process line.

.book Page 128 Monday, February 6, 2012 11:01 AM