beautypg.com

Service – KROHNE MFC 300 EN User Manual

Page 129

background image

SERVICE

7

129

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

7.2.5 OPTIMASS 8000k

The specified values are for guidance only.

• Driver = black and grey
• Sensor A = white and yellow
• Sensor B = violet and green
• Pt500 = red and violet (530…550 Ω) at ambient temperature
• Resistance values outside of the ranges specified above may indicate a circuit error. The

device may be in "start-up" mode or indicate measuring errors.

• All circuits should be isolated with respect to earth (device housing) and each other >20 MΩ.
• An exciter/sensor coil short circuit may cause the meter to be in "start-up" mode.

Size

Resistance (Ohm)

Driver

Sensor A/B

15

65

123

25

117

110

40

80

110

80

80

140

100

120

110

INFORMATION!
Failure of two or more of the upper circuits may indicate a measuring tube error. There may be
product in the housing. If this is the case, depressurise

depressurise

depressurise

depressurise the process line and immediately remove

the measuring device from the process line.

.book Page 129 Monday, February 6, 2012 11:01 AM