OPERATION

6

93

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

C6.2 Display

C6.2

Display

C6.2.1

Language

Language selection depends on the device version.

C6.2.2

Contrast

Adjust display contrast for extreme temperatures.

Setting: -9…0…+9

This change takes place immediately, not just when setting mode is exited!

C6.2.3

Default Display

Specification of the default display page that is returned to after a short

delay period.

Select: none (the current page is always active) / 1. meas. page (show this

page) / 2. meas. page (show this page) / status page (show only status

messages) / graphic page (trend display of the 1st measurement)

C6.2.4

Self Test

Not available at this time.

C6.2.5

SW.REV.UIS

Serial no. of the circuit board, version no. of user software, production date

of the circuit board

C6.3 and C6.4 1st Meas. Page and 2nd Meas. Page

C6.3

1st Meas. Page

…

stands for 3 = meas. page 1 and 4 = meas. page 2

C6.4

2nd Meas. Page

C6.….1

Function

Specify number of measured value lines (font size)

Select: one line / two lines / three lines

C6.….2

1st Line Variable

Specify 1st Line Variable.

Select measurement: Volume Flow / Mass Flow / Temperature / Density /

Velocity / Diagnosis 1 / Diagnosis 2 /

Depending on the settings for the concentration measurement, the following

measurements are possible:

Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 /

Concentration Flow 2

C6.….3

Range

0…100% of the measurement set in Fct. C5.….2

0…xx.xx _ _ _ (format and unit depend on the measurement)

C6.….4

Limitation

Limitation before applying the time constant.

xxx%; Range: -150…+150%

C6.….5

Low Flow Cutoff

Sets output to "0": x.xxx ± x.xxx %; range: 0.0…20%

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

C6.….6

Time Constant

Range: 000.1…100 s

C6.….7

1st Line Format

Specify decimal places.

Select: automatic (adaptation is automatic) /

X (= none) …X.XXXXXXXX (max. 8 digits) depends on size of font

C6.….8

2nd Line Variable

Specify 2nd Line Variable (only available if this 2nd line is activated)

Select: Bar Graph (for measurement selected in the 1st line ) / Volume

Flow / Mass Flow / Temperature / Density / Velocity / Bar Graph / Totaliser

1 / Totaliser 2 / Totaliser 3 / Operating Hours / Diagnosis 1 / Diagnosis 2

Depending on the settings for the concentration measurement, the following

measurements are possible:

Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 /

Concentration Flow 2

C6.….9

2nd Line Format

Specify decimal places.

Select: automatic (adaptation is automatic) /

X (= none) …X.XXXXXXXX (max. 8 digits) depends on size of font