C6.2 Display
C6.2
Display
C6.2.1
Language
Language selection depends on the device version.
C6.2.2
Contrast
Adjust display contrast for extreme temperatures.
Setting: -9…0…+9
This change takes place immediately, not just when setting mode is exited!
C6.2.3
Default Display
Specification of the default display page that is returned to after a short
delay period.
Select: none (the current page is always active) / 1. meas. page (show this
page) / 2. meas. page (show this page) / status page (show only status
messages) / graphic page (trend display of the 1st measurement)
C6.2.4
Self Test
Not available at this time.
C6.2.5
SW.REV.UIS
Serial no. of the circuit board, version no. of user software, production date
of the circuit board
C6.3 and C6.4 1st Meas. Page and 2nd Meas. Page
C6.3
1st Meas. Page
stands for 3 = meas. page 1 and 4 = meas. page 2
C6.4
2nd Meas. Page
C6. .1
Function
Specify number of measured value lines (font size)
Select: one line / two lines / three lines
C6. .2
1st Line Variable
Specify 1st Line Variable.
Select measurement: Volume Flow / Mass Flow / Temperature / Density /
Velocity / Diagnosis 1 / Diagnosis 2 /
Depending on the settings for the concentration measurement, the following
measurements are possible:
Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 /
Concentration Flow 2
C6. .3
Range
0…100% of the measurement set in Fct. C5. .2
0…xx.xx _ _ _ (format and unit depend on the measurement)
C6. .4
Limitation
Limitation before applying the time constant.
xxx%; Range: -150…+150%
C6. .5
Low Flow Cutoff
Sets output to "0": x.xxx ± x.xxx %; range: 0.0…20%
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value ≤ 1st value
C6. .6
Time Constant
Range: 000.1…100 s
C6. .7
1st Line Format
Specify decimal places.
Select: automatic (adaptation is automatic) /
X (= none) …X.XXXXXXXX (max. 8 digits) depends on size of font
C6. .8
2nd Line Variable
Specify 2nd Line Variable (only available if this 2nd line is activated)
Select: Bar Graph (for measurement selected in the 1st line ) / Volume
Flow / Mass Flow / Temperature / Density / Velocity / Bar Graph / Totaliser
1 / Totaliser 2 / Totaliser 3 / Operating Hours / Diagnosis 1 / Diagnosis 2
Depending on the settings for the concentration measurement, the following
measurements are possible:
Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 /
Concentration Flow 2
C6. .9
2nd Line Format
Specify decimal places.
Select: automatic (adaptation is automatic) /
X (= none) …X.XXXXXXXX (max. 8 digits) depends on size of font