OPERATION

6

85

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

C3.… Current Output X

C3.…

Current Output X

X stands for one of the connection terminals A, B or C

…

stands for Fct. no. C3.2 (A) / C3.3 (B) / C3.4 (C)

C3.….1

Range 0%…100%

HART

®

current output: 4…20 mA

Current range for the selected measurement, e.g. 4...20 mA,

corresponds to 0…100%

Note: with a 0...20 mA current output, HART in Fct. C6.8.1 must be shut off!

xx.x … xx.x mA; range: 0.00…20 mA

(condition: 0 mA ≤ 1st value ≤ 2nd value ≤ 20 mA)

C3.….2

Extended Range

Defines the min. and max. limits.

xx.x … xx.x mA; range: 03.5…21.5 mA

(condition: 0 mA ≤ 1st value ≤ 2nd value ≤ 21.5 mA)

C3.….3

Error Current

Specify error current.

xx.x mA; range: 3…22 mA (condition: outside of extended range)

C3.….4

Error Condition

The following error conditions can be selected.

Select: error in device (error category [F]) / application error (error category

[F]) / out of specification (error category [S])

C3.….5

Measurement

Measurements for activating the output.

Select measurement: Volume Flow / Mass Flow / Temperature / Density /

Velocity / Diagnosis 1 / Diagnosis 2 /

Depending on the settings for the concentration measurement, the following

measurements are possible:

Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 /

Concentration Flow 2

C3.….6

Range

0…100% of the measurement set in Fct. C3.….5

0…xx.xx _ _ _ (format and unit depend on the measurement, see above)

C3.….7

Polarity

Set polarity, please note Flow Direction in C1.3.1!

Select: both polarities (plus and minus values are displayed) / positive

polarity (display for negative values = 0) / negative polarity (display for

positive values = 0) / absolute value (use for the output)

C3.….8

Limitation

Limitation before applying the time constant.

±

xxx … ±xxx%; range: -150…+150%

C3.….9

Low Flow Cutoff

Sets output value to "0"

x.xxx ± x.xxx%; range: 0.0…20%

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

C3.….10

Time Constant

Range: 000.1…100 s

C3.….11

Special Function

Automatic Range; select:

Off (switched off)

Automatic range (range is changed automatically, extended lower range,

only makes sense together with a status output)

External range (change by control input, extended lower range, control input

must also be activated)

.book Page 85 Monday, February 6, 2012 11:01 AM