9

DESCRIPTION OF HART INTERFACE

152

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

9.4 Inputs/outputs and HART

®

dynamic variables and device variables

The signal converter is available with various input/output combinations.

The connection of the terminals A…D to the HART

®

dynamic variables PV, SV, TV and 4V depends

on the device version.

PV = Primary Variable; SV = Secondary Variable; TV = Third Variable; 4V = Fourth Variable

The signal converter can provide up to 14 measured values. The measured values are accessible
as so-called HART

®

device variables and can be connected to the HART

®

dynamic variables. The

availability of these variables depends on the device versions and the settings.

Code = device variable code

Device variables

Signal converter version

HART

®

dynamic variable

PV

SV

TV

4V

Basic I/O, connection terminals

A

D

-

-

Modular I/O and Ex i I/O, connection terminals

C

D

A

B

HART

®

device variable

Code

Type

Explanations

Flow Velocity

20

linear

Volume Flow

21

linear

Mass Flow

22

linear

Temperature

23

linear

Density

24

linear

Concentration 1

25

linear

Only when concentration measurement

available and concentration function 1

not switched off.

Concentration 2 / Diagnosis 3

26

linear

Concentration 2

Concentration 2

Concentration 2

Concentration 2: only when

concentration measurement available

and concentration function 2 not

switched off.

Diagnosis 3

Diagnosis 3

Diagnosis 3

Diagnosis 3: only available when

diagnosis value 3 not switched off and

concentration function 2 switched off.

Concentration Flow 1

27

linear

Only available when concentration

measurement switched on and

concentration function 1 not switched

off.

Concentration Flow 2

28

linear

Only available when concentration

measurement switched on and

concentration function 2 not switched

off.

Diagnosis 1

29

linear

Available when diagnosis value 1 not

switched off.

.book Page 152 Monday, February 6, 2012 11:01 AM