Operation – KROHNE MFC 300 EN User Manual
Page 81
OPERATION
6
81
MFC 300
www.krohne.com
02/2012 - 4000498602 - MA MFC 300 R03 en
6.4.2 Menu B, Test
No.
Function
Settings / descriptions
B1 Simulation
B1
Simulation
Displayed values are simulated.
B1.1
Mass Flow
Simulation of mass flow
Set value (range and units depend on measurement)
break (exit function without simulation)
Query: start simulation?
Settings: no (exit function without simulation) / yes (start simulation)
B1.2
Density
Sequence and settings similar to B1.1, see above!
X stands for one of the connection terminals A, B, C or D
stands for Fct. B1.4...1.7
B1.3
Temperature
B1.
Current Output X
Simulation X
X stands for one of the connection terminals A, B, C or D
Sequence and settings similar to B1.1, see above!
For pulse output a set number of pulses are output in 1 s!
B1.
Pulse Output X
B1.
Frequency Output X
B1.
Control Input X
B1.
Limit Switch X
B1.
Status Output X
B2 Actual Values
B2
Actual Values
Display of current values;
Exit the displayed function by pressing the ^ key.
B2.1
Operating Hours
Device operating hours
B2.2
Mass Flow
Current unfiltered mass flow
B2.3
Volume flow
Current unfiltered volume flow
B2.4
Velocity
Current unfiltered flow velocity
B2.5
Density
Current unfiltered density
B2.6
Temperature
Current unfiltered temperature
B2.7
Strain MT
Current value for the measuring tube strain
B2.8
Strain IC
Current value for the inner cylinder strain
B2.9
Tube Frequency
Current vibration frequency of the measuring tube
B2.10
Drive Level
Current drive level to activate vibration
B2.11
Sensor A Level
Current vibration amplitude
B2.12
Sensor B Level
B2.13
2 Phase Signal
2 phase indicator value
B2.14
SE PCB Temperature
Temperature of sensor electronics
B2.15
BE PCB Temperature
Temperature of signal converter
B2.16
Act. Operat. Mode
Current operating mode