Operation – KROHNE MFC 300 EN User Manual

Page 81

background image

OPERATION

6

81

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

6.4.2 Menu B, Test

No.

Function

Settings / descriptions

B1 Simulation

B1

Simulation

Displayed values are simulated.

B1.1

Mass Flow

Simulation of mass flow

Set value (range and units depend on measurement)

break (exit function without simulation)

Query: start simulation?

Settings: no (exit function without simulation) / yes (start simulation)

B1.2

Density

Sequence and settings similar to B1.1, see above!

X stands for one of the connection terminals A, B, C or D

…

stands for Fct. B1.4...1.7

B1.3

Temperature

B1.…

Current Output X

Simulation X

X stands for one of the connection terminals A, B, C or D

Sequence and settings similar to B1.1, see above!

For pulse output a set number of pulses are output in 1 s!

B1.…

Pulse Output X

B1.…

Frequency Output X

B1.…

Control Input X

B1.…

Limit Switch X

B1.…

Status Output X

B2 Actual Values

B2

Actual Values

Display of current values;

Exit the displayed function by pressing the ^ key.

B2.1

Operating Hours

Device operating hours

B2.2

Mass Flow

Current unfiltered mass flow

B2.3

Volume flow

Current unfiltered volume flow

B2.4

Velocity

Current unfiltered flow velocity

B2.5

Density

Current unfiltered density

B2.6

Temperature

Current unfiltered temperature

B2.7

Strain MT

Current value for the measuring tube strain

B2.8

Strain IC

Current value for the inner cylinder strain

B2.9

Tube Frequency

Current vibration frequency of the measuring tube

B2.10

Drive Level

Current drive level to activate vibration

B2.11

Sensor A Level

Current vibration amplitude

B2.12

Sensor B Level

B2.13

2 Phase Signal

2 phase indicator value

B2.14

SE PCB Temperature

Temperature of sensor electronics

B2.15

BE PCB Temperature

Temperature of signal converter

B2.16

Act. Operat. Mode

Current operating mode

.book Page 81 Monday, February 6, 2012 11:01 AM