ELECTRICAL CONNECTIONS

4

65

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

Control input passive, Ex i I/Os

• 5.5 V ≤ U

ext

≤ 32 VDC

• I

max

= 6 mA at U

ext

≤ 24 V

I

max

= 6.5 mA at U

ext

≤ 32 V

• Switching point for identifying "contact open or closed":

Contact open (off): U

0

≤ 3.5 V with I ≤ 0.5 mA

Contact closed (on): U

0

≥ 5.5 V with I ≥ 4 mA

• X designates the connection terminals B, if available.

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.

INFORMATION!

Any connection polarity.

Figure 4-32: Control input passive C

p

Exi

1 Signal