Technical data – KROHNE MFC 300 EN User Manual
Page 137
TECHNICAL DATA
8
137
MFC 300
www.krohne.com
02/2012 - 4000498602 - MA MFC 300 R03 en
Materials
Signal converter housing
Standard
Standard
Standard
Standard
Versions C and F: die-cast aluminium (polyurethane coated)
Version W: polyamide-polycarbonate
Version R: aluminium, stainless steel and aluminium sheet, partially polyester-
coated
Option
Option
Option
Option
Versions C and F: stainless steel 316 L (1.4408)
Measuring sensor
For housing material, process connections, measuring tubes, accessories and
gaskets, refer to technical data for the measuring sensor.
Electrical connection
General
Electrical connection is carried out in conformity with the VDE 0100 directive
"Regulations for electrical power installations with line voltages up to 1000 V" or
equivalent national regulations.
Power supply
Standard: 100…230 VAC (-15% / +10%), 50/60 Hz
Option 1: 24 VDC (-55% / +30%)
Option 2: 24 VAC/DC (AC: -15% / +10%, 50/60 Hz; DC: -25% / +30%)
Power consumption
AC: 22 VA
DC: 12 W
Signal cable
Only for remote versions.
4 core shielded cable. Detailed specifications available on request.
Length: max. 300 m / 1000 ft
Cable entries
Standard: M20 x 1.5 (8...12 mm)
Option: ½" NPT, PF ½
.book Page 137 Monday, February 6, 2012 11:01 AM