C3.… Status Output X

C3.…

Status Output X

X (Y) stands for one of the connection terminals A, B, C or D

…

stands for Fct. no. C3.2 (A) / C3.3 (B) / C3.4 (C) / C3.5 (D)

C3.….1

Mode

The output shows the following measuring conditions:

out of specification (output activated, signals application error or error in

device refer to

Status messages and diagnostic information on page 113 /

application error (output activated, signals application error or error in

device refer to

Status messages and diagnostic information on page 113 /

polarity flow (polarity of the current flow) /

over range flow (over range of the flow) /

totaliser 1 preset (activates when totaliser X preset value is reached) /

totaliser 2 preset (activates when totaliser X preset value is reached) /

totaliser 3 preset (activates when totaliser X preset value is reached) /

output A (activated by the status of output Y, additional output data see

below) /

output B (activated by the status of output Y, additional output data see

below) /

output C (activated by the status of output Y, additional output data see

below) /

output D (activated by the status of output Y, additional output data see

below) /

off (switched off) /

empty pipe (when pipe empty, output activated) /

Error in device (when error, output activated)

C3.….2

Current Output Y

Only appears if output A...C is set under "mode" (see above), and this output

is a "current output".

Polarity (is signalled)

Over Range (is signalled)

Automatic Range signals lower range

C3.….2

Frequency Output Y and

Pulse Output Y

Only appears if output A, B or D is set under "mode" (see above), and this

output is a "frequency/pulse output".

Polarity (is signalled)

Over Range (is signalled)

C3.….2

Status Output Y

Only appears if output A...D is set under "mode" (see above), and this output

is a "status output".

Same signal (like other connected status output, signal can be inverted, see

below)

C3.….2

Limit Switch Y and Control

Input Y

Only appears if output A...D / input A or B is set under "mode" (see above),

and this output / input is a "limit switch / control input".

Status off (is always selected here if status output X is connected with a limit

switch / control input Y.

C3.….2

Off

Only appears if output A...D is set under "mode" (see above) and this output

is switched off.

C3.….3

Invert Signal

Off (activated output supplies a high current, switch closed)

On (activated output supplies a low current, switch open)

C3.….4

Information

Serial no. of the I/O board, software version no. and production date of the

circuit board

C3.….5

Simulation

Sequence see B1.… status output X