OPERATION
6
119
MFC 300
www.krohne.com
02/2012 - 4000498602 - MA MFC 300 R03 en
Typical frequency values (at 20°C / 68°F)
Problems with zero calibration
1 Stop the flow.
2 Set the counter in Fct. C3.y.1 to totaliser.
3 Set the low flow cutoff in Fct. C3.y.3 to zero.
4 Perform automatic zero calibration.
5 Reset totaliser and count for 2 minutes.
6 Compare the added value to the specified zero calibration stability.
Possible causes of poor zero calibration:
• Valves not completely closed, air or gas bubbles or deposits on measuring tube
Size
Titanium
Stainless steel
Hastelloy
®
Tantalum
Empty
Water
Empty
Water
Empty
Water
Empty
Water
1000 - 15
438±10
412±10
1000 - 25
605±20
523±20
1000 - 40
494±10
414±10
1000 - 50
583±10
453±10
2000 - 100
341±6
267±6
2000 - 150
330±6
259±6
2000 - 250
299±6
227±6
3000 - 01
258±6
251±6
266±6
258±6
3000 - 03
320±6
310±6
320±6
310±6
3000 - 04
455±6
435±6
455±6
435±6
7000 - 06
316±10
301±10
374±10
362±10
7000 - 10
406±10
371±10
441±10
417±10
439±10
416±10
348±10
330±10
7000 - 15
502±10
432±10
578±10
519±10
566±10
509±10
430±10
394±10
7000 - 25
614±10
483±10
692±10
580±10
687±10
581±10
515±10
449±10
7000 - 40
462±10
367±10
558±10
467±10
556±10
468±10
417±10
360±10
7000 - 50
488±10
357±10
514±10
418±10
539±10
431±10
403±10
333±10
7000 - 80
480±10
338±10
490±10
370±10
493±10
381±10
8000 - 15
226±3
202±3
8000 - 25
280±3
242±3
8000 - 40
271±3
238±3
8000 - 80
241±3
22±3
8000 - 100
264±3
228±3
INFORMATION!
For best results, perform zero calibration with the medium at operating temperature.