Typical frequency values (at 20°C / 68°F)

Problems with zero calibration

1 Stop the flow.
2 Set the counter in Fct. C3.y.1 to totaliser.
3 Set the low flow cutoff in Fct. C3.y.3 to zero.
4 Perform automatic zero calibration.
5 Reset totaliser and count for 2 minutes.
6 Compare the added value to the specified zero calibration stability.

Possible causes of poor zero calibration:

• Valves not completely closed, air or gas bubbles or deposits on measuring tube

Size

Titanium

Stainless steel

Hastelloy

®

Tantalum

Empty

Water

Empty

Water

Empty

Water

Empty

Water

1000 - 15

438±10

412±10

1000 - 25

605±20

523±20

1000 - 40

494±10

414±10

1000 - 50

583±10

453±10

2000 - 100

341±6

267±6

2000 - 150

330±6

259±6

2000 - 250

299±6

227±6

3000 - 01

258±6

251±6

266±6

258±6

3000 - 03

320±6

310±6

320±6

310±6

3000 - 04

455±6

435±6

455±6

435±6

7000 - 06

316±10

301±10

374±10

362±10

7000 - 10

406±10

371±10

441±10

417±10

439±10

416±10

348±10

330±10

7000 - 15

502±10

432±10

578±10

519±10

566±10

509±10

430±10

394±10

7000 - 25

614±10

483±10

692±10

580±10

687±10

581±10

515±10

449±10

7000 - 40

462±10

367±10

558±10

467±10

556±10

468±10

417±10

360±10

7000 - 50

488±10

357±10

514±10

418±10

539±10

431±10

403±10

333±10

7000 - 80

480±10

338±10

490±10

370±10

493±10

381±10

8000 - 15

226±3

202±3

8000 - 25

280±3

242±3

8000 - 40

271±3

238±3

8000 - 80

241±3

22±3

8000 - 100

264±3

228±3

INFORMATION!
For best results, perform zero calibration with the medium at operating temperature.