B) Manual calibration
Under certain conditions, zero calibration is not possible and will be cancelled:
• The product is still flowing. The shut-off valves are not closed tightly enough.
• There are still gas bubbles in the fluid.
Remedy: Flush sensor and repeat calibration
With some media, it may be difficult to perform zero calibration. In such cases, there are various
methods to still achieve good zero calibration:
Key
Display
Description and setting
>
A
Quick Setup
Press and hold for 2.5 s, then release
the key.
2 x ↓
C
Setup
3 x >
C1.1.1
Zero Calibration
>
Calibrate Zero?
Break
3 x ↓
Calibrate Zero?
Manual
Zero Calibration
+XX.XXX%
Display currently stored zero
calibration in %.
(Warning, value can be changed!)
Possible manual entry of zero
calibration.
Storing the displayed zero
calibration.
5 x ^
Save Configuration?
Yes
^
Display page
Medium
Possible solutions
Media that tend to evaporate or outgas
Increase pressure.
Two-phase media (slurries), that contain solids that
may fall out.
Only fill the sensor with the carrier medium.
Two-phase media in which the solids or gas-
forming components cannot be separated.
Fill sensor with another liquid, e.g. water.