OPERATION

6

115

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

Measurements out of specification

F SE Comms. Failure

Communication error with sensor

electronics (SE). No measurement data

available.

Check wiring and grounding. Replace

electronics.

F Active Settings

Error during the CRC check of the active

settings.

Upload backup 1 or backup 2 settings, check

and adjust if necessary.

F Factory Settings

Error during the CRC check of the factory

settings.

F Backup 1 Settings

Error during the CRC check of the backup 1

or 2 settings.

Save active settings in backup 1 or 2.

F Backup 2 Settings

Messages on the display Description

Actions

Status: S _ _ _ _ _

Out of specification, measurement
continues, accuracy possibly less.

Maintenance required.

S Out Of Specification

Device maintenance necessary; measured

values only conditionally usable.

Group message, when errors as described

below or other influences occur.

S Overflow Totaliser 1

This is totaliser 1 or FB2 (with Profibus).

Totaliser has overrun and started again at

zero.

Check totaliser format.

S Overflow Totaliser 2

This is totaliser 2 or FB3 (with Profibus).

Totaliser has overrun and started again at

zero.

S Overflow Totaliser 3

This is totaliser 3 or FB4 (with Profibus). Not

available without IO2. Totaliser has overrun

and started again at zero.

S Backplane Invalid

The data record on the backplane is invalid.

The CRC check has revealed a fault.

No data can be loaded from the backplane

when replacing electronics. Save the data to

the backplane again (Service).

S SE PCB Temperature

Temperature on SE PCB is exceeding

maximum limit.

Check process and ambient temperature.

Check wiring. Exchange the sensor

electronics (SE).

S Startup

The device is in startup mode. If the system

cannot start from start-up mode or came

into the startup mode from the

measurement function, the message "F

application error" also appears.

Check process conditions (air). Check device

settings C1.7.1...C1.7.3. Check sensor

resistances.

S Power Fail

For custody transfer applications. Indicates

a failure of power supply. No measurement

is possible during power failure.

S Tube Temperature

Process temperature is outside limits of

sensor. Failure of sensor may occur if

prolonged.

Check settings C1.7.4 and C1.7.5. Reduce

process temperature.

S Density

Process density is over range.

Check process conditions.

S Sensor Signal Error

DC component of measuring sensor signal

is too high.

Check sensor resistances. Replace sensor.

S Res. Circ. Defective

Pt500 sensor is defective. Temperature

measurement and compensation is

unreliable.

Check sensor resistances. Replace sensor.

S Sensor Levels

Tube vibration amplitude too low.

Check the process conditions (air).

Messages on the display Description

Actions

Status: F _ _ _ _ _

Application fault, device OK, but
measured values affected.

Application test or operator action
necessary.