OPERATION

6

87

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

C3.….8

Low Flow Cutoff

Sets output value to "0":

x.xxx ± x.xxx%; range: 0.0…20%

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

C3.….9

Time Constant

Range: 000.1…100 s

C3.….10

Invert Signal

Select: Off (activated output generates a high current at the output, switch

closed) /

On (activated output generates a low current at the output, switch open)

C3.….11

Phase Shift w.r.t. B

Only available when configuring the A or D terminal and only if output B is a

pulse or frequency output. If setting in Fct. C2.5.6 is "Both Polarities", the

phase shift is prefixed by a symbol, e.g. -90° and +90°.

Select: off (no phase shift) / 0° phase shift (between outputs A or D and B,

inversion possible) / 90° phase shift (between outputs A or D and B, inversion

possible) / 180° phase shift (between outputs A or D and B, inversion

possible)

C3.3.11

Special Functions

This function is only available at the terminal B frequency output. At the

same time, 2 frequency outputs must be available:

1st output at terminal A or D / 2nd output at terminal B

The B output is operated as a slave output, controlled and set using master

output A or D

Selection: off (no phase shift) / phase shift w.r.t. D or A (slave output is B and

master output is D or A)

C3.….12

Information

Serial no. of the I/O board, software version no. and production date of the

circuit board

C3.….13

Simulation

Sequence see B1.… Frequency Output X

C3.… Pulse Output X

C3.…

Pulse Output X

X stands for one of the connection terminals A, B or D

…

stands for Fct. no. C3.2 (A) / C3.3 (B) / C3.5 (D)

C3.….1

Pulse Shape

Specify the pulse shape.

Select: Symmetric (about 50% on and 50% off) / Automatic (constant pulse

with about 50% on and 50% off at 100% pulse rate) / Fixed (fixed pulse rate,

setting see below Fct. C3.….3 100% pulse rate)

C3.….2

Pulse Width

Only available if set to "fixed" in Fct. C3.….1

Range: 0.05…2000 ms

Note: max. setting value Tp [ms] ≤ 500 / max. pulse rate [1/s], gives the pulse

width = time where the output is activated

C3.….3

Max. Pulse Rate

Pulse rate for 100% of the measuring range

Range: 0.0…10000 1/s

Limitation 100% pulse rate ≤ 100/s: I

max

≤ 100 mA

Limitation 100% pulse rate > 100/s: I

max

≤ 20 mA

C3.….4

Measurement

Measurements for activating the output.

Select: Volume Flow / Mass Flow

C3.….5

Pulse Value Unit

Selection of the unit from a list, depending on the measurement.

C3.….6

Value p. Pulse

Set value for volume or mass per pulse.

xxx.xxx, range in [l] or [kg] (volume or mass for current output C3.….6)

At max. pulse rate see above 3.….3 Pulse Output.

C3.….7

Polarity

Set polarity, please note Flow Direction in C1.3.2!

Select: both polarities (plus and minus values are displayed) / positive

polarity (display for negative values = 0) / negative polarity (display for

positive values = 0) / absolute value (use for the output)