background image

4

ELECTRICAL CONNECTIONS

48

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

4.9.3 19" rack-mounted housing (28 TE), electrical connection of the inputs and outputs

• For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from

electrical interferences (EMC).

• Terminal A+ is only operable in the basic version.

• Connect the conductor to the multipolar plug according to the illustration.
• The signal cable shield is connected to the Pin S.
• Press the plug into the connector.

4.9.4 Laying electrical cables correctly

1 Lay the cable in a loop just before the housing.
2 Tighten the screw connection of the cable entry securely.
3 Never mount the housing with the cable entries facing upwards.
4 Seal cable entries that are not needed with a plug.

DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!

Figure 4-11: Terminal compartment for inputs and outputs in rack-mounted housing

1 Shielding

Figure 4-12: Protect housing from dust and water