Technical data – KROHNE MFC 300 EN User Manual

Page 135

background image

TECHNICAL DATA

8

135

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

Display and user interface

Display and user interface

Display and user interface

Display and user interface

Graphic display

LC display, backlit white.

Size: 128x64 pixels, corresponds to 59x31 mm = 2.32"x1.22"

Display turnable in 90° steps.

Ambient temperatures below -25°C / -13°F may affect the readability of the display.

Operating elements

4 optical keys for operator control of the signal converter without opening the

housing.

Infrared interface for reading and writing all parameters with IR interface (option)

without opening the housing.

Remote control

PACTware

®

(including Device Type Manager (DTM))

HART

®

Hand Held Communicator from Emerson Process

AMS

®

from Emerson Process

PDM

®

from Siemens

All DTMs and drivers are available free of charge from the manufacturer's website.

Display functions

Display functions

Display functions

Display functions

Operating menu

Setting the parameters using 2 measured value pages, 1 status page, 1 graphics

page (measured values and graphics are freely adjustable)

Language display texts (as

language package)

Standard: English, French, German, Dutch, Portuguese, Swedish, Spanish, Italian

Eastern Europe (in preparation): English, Slovenian, Czech, Hungarian

Northern Europe (in preparation): English, Danish, Polish

China (in preparation): English, Chinese

Russia: English, Russian

Measurement functions

Units:

Units:

Units:

Units: Metric, British and US units selectable as desired from lists for volume/mass

flow and counting, velocity, temperature, pressure

Measured values:

Measured values:

Measured values:

Measured values: Mass flow, total mass, temperature, density, volume flow, total

volume, velocity, flow direction (not displayed unit – but available via outputs), BRIX,

Baume, NaOH, Plato, API, mass concentration, volume concentration

Diagnostic functions

Standards:

Standards:

Standards:

Standards: according to VDI / NAMUR / WIB 2650 (pending) and functions going

beyond that

Status messages:

Status messages:

Status messages:

Status messages: Output of status messages optional via display, current and/or

status output, HART® or bus interface

Sensor diagnostics:

Sensor diagnostics:

Sensor diagnostics:

Sensor diagnostics: Sensor values, drive level, measuring tube frequency, MT

(measuring tube) strain, IC (inner cylinder) strain, sensor electronics/board

electronics temperature, 2-phase flow signal

Measurement accuracy

Reference conditions

Medium: water

Temperature: 20°C / 68°F

Pressure: 1 bar / 14.5 psi

Maximum measuring error

±

0.10% of the measured value ± zero point stability (depending on the measuring

sensor)

Current output electronics: ±5 µA

Repeatability

±

0.05% ± zero point stability (depending on the measuring sensor)

.book Page 135 Monday, February 6, 2012 11:01 AM