beautypg.com

Operation – KROHNE MFC 300 EN User Manual

Page 86

background image

6

OPERATION

86

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

C3.….12

Threshold

Appears only when Fct. C3.….11 threshold is activated between extended

and normal range. The automatic range function always changes from the

extended to the normal range when the 100% current is reached.

The upper 100% value of the hysteresis is then = 0. The threshold is then the

hysteresis value, instead of "threshold ± hysteresis" as shown in the display.

Range: 5.0…80%

(1st value = switching point / 2nd value = hysteresis),

condition: 2nd value ≤ 1st value

C3.….13

Information

Serial no. of the I/O board, software version no. and production date of the

circuit board

C3.….14

Simulation

Sequence see B1.… Current Output X

C3.….15

4mA Trimming

Trimming of the current at 4 mA

Reset to 4 mA restores the factory calibration.

Used for HART

®

setting.

C3.….16

20mA Trimming

Trimming of the current at 20 mA

Reset to 20 mA restores the factory calibration.

Used for HART

®

setting.

C3.… Frequency Output X

C3.…

Frequency Output X

X stands for one of the connection terminals A, B or D

…

stands for Fct. no. C3.2 (A) / C3.3 (B) / C3.5 (D)

C3.….1

Pulse Shape

Specify the pulse shape.

Select: symmetric (about 50% on and 50% off) / automatic (constant pulse

with about 50% on and 50% off at 100% pulse rate) / fixed (fixed pulse rate,

setting see below Fct. C3.….3 100% pulse rate)

C3.….2

Pulse Width

Only available if set to "fixed" in Fct. C3.….1.

Range: 0.05…2000 ms

Note: max. setting value Tp [ms] ≤ 500 / max. pulse rate [1/s], gives the pulse

width = time where the output is activated

C3.….3

100% Pulse Rate

Pulse rate for 100% of the measuring range

Range: 0.0…10000 1/s

Limitation 100% pulse rate ≤ 100/s: I

max

≤ 100 mA

Limitation 100% pulse rate > 100/s: I

max

≤ 20 mA

C3.….4

Measurement

Measurements for activating the output.

Select measurement: Volume Flow / Mass Flow / Temperature / Density /

Velocity / Diagnosis 1 / Diagnosis 2 /

Depending on the settings for the concentration measurement, the following

measurements are possible:

Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 /

Concentration Flow 2

C3.….5

Range

0…100% of the measurement set in Fct. C3.….4

0…xx.xx _ _ _ (format and unit depend on the measurement, see above)

C3.….6

Polarity

Set measured value polarity, please note flow direction in C1.3.2!

Select: both polarities (plus and minus values are displayed) / positive

polarity (display for negative values = 0) / negative polarity (display for

positive values = 0) / absolute value (use for the output)

C3.….7

Limitation

Limitation before applying the time constant.

±

xxx … ±xxx%; range: -150…+150%

.book Page 86 Monday, February 6, 2012 11:01 AM