Calibrating a touch trigger probe

17.3

17

TNC 620 | User's Manual Cycle Programming | 5/2013

403

17.3

Calibrating a touch trigger probe

In order to precisely specify the actual trigger point of a 3-D touch
probe, you must calibrate the touch probe, otherwise the TNC
cannot provide precise measuring results.

Always calibrate a touch probe in the following cases:

Commissioning

Stylus breakage

Stylus exchange

Change in the probe feed rate

Irregularities caused, for example, when the
machine heats up

Change of active tool axis

The TNC assumes the calibration values for the
active probe system directly after the calibration
process. The updated tool data become effective
immediately, and a new tool call is not necessary.

During calibration, the TNC finds the "effective" length of the stylus
and the "effective" radius of the ball tip. To calibrate the 3-D touch
probe, clamp a ring gauge or a stud of known height and known
radius to the machine table.

The TNC provides calibration cycles for calibrating the length and
the radius:

Press the TOUCH PROBE soft key

Display the calibration cycles: Press CALIBRATE
TS

Select the calibration cycle

Calibration cycles of the TNC

Soft key

Function

Page

Calibrating the length

407

Measure the radius and the center
offset using a calibration ring

408

Measure the radius and the center
offset using a stud or a calibration
pin

410

Measure the radius and the center
offset using a calibration sphere

405