Error codes – AMETEK ReFlex Programming Manual User Manual
Page 95

AC Module Remote Programming
ReFlex Power™ Programming Manual
M380056-03 Rev L
91
4.4
ERROR CODES
The error codes exist in the range of [-32768, 32767]. SCPI reserves the
negative error codes and zero, while the error codes greater than zero and in the
range of -399 to -300 are defined as device-specific. The negative error codes in
the range of -499 to -100 are standardized errors that are reported in the
Standard Event Status Register and the Error/Event Queue. Issue the
SYSTem:ERRor?
query to read the error codes, which are stored in the
Error/Event Queue.
4.4.1
SCPI
E
RROR
C
ODES
The following error codes are defined by the SCPI Standard 1999.0, and are
supported by the ReFlex Power™ ACPS:
0, No error
The error queue is empty.
-102, Syntax error
An unrecognized command or data type was encountered.
-200, Execution error
This is a generic error when a specific error could not be identified. An execution
error could be the result of the following condition:
A command could not be executed due to the current operational condition.
-222, Range Error
A
when the maximum allowed value is 135V.
-350, Queue overflow
The Error/Event Queue can contain up to 10 entries. If more than 10 error/event
conditions are logged before the SYSTem:ERRor? query, an overflow will occur;
the last queue entry will be overwritten with error -350. When the queue
overflows, the least recent error/event remains in the queue and the most recent
error/event is discarded.
NOTE: Some SCPI commands will generate replies as follows, which can be
viewed with the SYST:ERR? command: