Measure scpi commands – AMETEK ReFlex Programming Manual User Manual
Page 63

DC Module Remote Programming
ReFlex Power™ Programming Manual
M380056-03 Rev L
59
Command
Description
:FIVEPOINT?
Returns 5 pairs of voltage calibration setpoint count and
related voltage values.
:PROTection
:COUNTS
Set Protection value 0-4095
:MODule
:VOLTage?
Returns Module Voltage rating
:CURRent?
Returns Module Current rating
:LOCK
Disables access to the non-volatile memory. Prevents
attempts to store calibration values.
:UNLock
Unlocks the non-volatile memory to store calibration
constants. The access string is “6867”.
:STORe
Stores the calibration constants in non-volatile memory.
:REVERT:FACTory
Restores the original factory calibration, into non-volatile
memory. Wait 30 seconds and then Power cycle.
Note 1: If CAL:INIT:STATe is ‘1’, the module will automatically turn on as if an OUTP:STATe 1
command had been received. This will only occur during initial power up, *RST will only restore
the saved CAL:INIT values. The unit will not automatically turn on, if enabled, after a *RST.
3.7
MEASURE SCPI COMMANDS
3.7.1
MEASURE
SCPI
C
OMMAND
S
UMMARY
MEASure[n]
:CURRent?
:POLARity?
:VOLTage?
:POWer?
3.7.2
MEASURE
SCPI
C
OMMAND
R
EFERENCE
Command
Description
MEASure
Measure subsystem. Module ‘n’.
:CURRent?
Returns the floating point value of the DC output current in amps.
The value will be negative if the polarity relay is in the INV state.
:POLARity?
Returns Polarity Relay setting.
:VOLTage?
Returns the floating point value of the DC output voltage in volts.
The value will be negative if the polarity relay is in the INV state.
:POWer?
Returns the power in Watts that are delivered to the Load.